Background: A new dual resolution imaging x-ray detector system (Canon Medical Systems Corporation, Tochigi, Japan) has a standard resolution 194 µm pixel conventional flat-panel detector (FPD) mode and a high-resolution 76 µm high-definition (Hi-Def) mode in a single unit. The Hi-Def mode enhances the visualization of the intravascular devices.
Objective: We report the clinical experience and physician evaluation of this new detector system with Hi-Def mode for the treatment of intracranial aneurysms using a Pipeline embolization device (PED).
Methods: During intervention at our institute, under large field of view (FOV) regular resolution FPD mode imaging, the catheter systems and devices were first guided to the proximity of the treatment area. Final placement and deployment of the PED was performed under Hi-Def mode guidance. A post-procedure 9-question physician survey was conducted to qualitatively assess the impact of Hi-Def mode visualization on physicians' intraoperative decision-making. One-sample t-test was performed on the responses from the survey. Dose values reported by the x-ray unit were also recorded.
Results: Twenty-five cases were included in our study. The survey results indicated that, for each of the nine questions, the physicians in all cases indicated that the Hi-Def mode improved visualization compared with the FPD mode. For the 25 cases, the mean cumulative entrance air kerma was 2.35 Gy, the mean dose area product (DAP) was 173.71 Gy.cm, and the mean x-ray exposure time was 39.30 min.
Conclusions: The Hi-Def mode improves visualization of flow diverters and may help in achieving more accurate placement and deployment of devices.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1136/neurintsurg-2019-015551 | DOI Listing |
J Neurointerv Surg
September 2020
Canon Stroke and Vascular Research Center, UB Neurosurgery, University at Buffalo, State University of New York, Buffalo, New York, USA
Background: A new dual resolution imaging x-ray detector system (Canon Medical Systems Corporation, Tochigi, Japan) has a standard resolution 194 µm pixel conventional flat-panel detector (FPD) mode and a high-resolution 76 µm high-definition (Hi-Def) mode in a single unit. The Hi-Def mode enhances the visualization of the intravascular devices.
Objective: We report the clinical experience and physician evaluation of this new detector system with Hi-Def mode for the treatment of intracranial aneurysms using a Pipeline embolization device (PED).
J Neuroimaging
September 2019
Canon Stroke and Vascular Research Center, University at Buffalo, Buffalo, NY.
Background And Purpose: Visualization of structural details of treatment devices during neurointerventional procedures can be challenging. A new true two-resolution imaging X-ray detector system features a 194 µm pixel conventional flat-panel detector (FPD) mode and a 76 µm pixel high-resolution high-definition (Hi-Def) zoom mode in one detector panel. The Hi-Def zoom mode was developed for use in interventional procedures requiring superior image quality over a small field of view (FOV).
View Article and Find Full Text PDF