Background And Purpose: Visualization of structural details of treatment devices during neurointerventional procedures can be challenging. A new true two-resolution imaging X-ray detector system features a 194 µm pixel conventional flat-panel detector (FPD) mode and a 76 µm pixel high-resolution high-definition (Hi-Def) zoom mode in one detector panel. The Hi-Def zoom mode was developed for use in interventional procedures requiring superior image quality over a small field of view (FOV). We report successful use of this imaging system during intracranial aneurysm treatment in 1 patient with a Pipeline-embolization device and 1 patient with a low-profile visualized intramural support (LVIS Blue) device plus adjunctive coiling.
Methods: A guide catheter was advanced from the femoral artery insertion site to the proximity of each lesion using standard FPD mode. Under magnified small FOV Hi-Def imaging mode, an intermediate catheter and microcatheters were guided to the treatment site, and the PED and LVIS Blue plus coils were deployed. Radiation doses were tracked intraprocedurally.
Results: Critical details, including structural changes in the PED and LVIS Blue and position and movement of the microcatheter tip within the coil mass, were more readily apparent in Hi-Def mode. Skin-dose mapping indicated that Hi-Def mode limited radiation exposure to the smaller FOV of the treatment area.
Conclusions: Visualization of device structures was much improved in the high-resolution Hi-Def mode, leading to easier, more controlled deployment of stents and coils than conventional FPD mode.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7150547 | PMC |
http://dx.doi.org/10.1111/jon.12652 | DOI Listing |
IEEE Trans Biomed Circuits Syst
May 2025
In the practical implementations of wearable sensors, motion artifacts with large amplitudes often cause signal chain saturation, significantly degrading biopotential signal integrity. Similarly, rapid stimulation artifacts are inevitable during closed-loop brain stimulation therapy, posing additional challenges for real-time signal acquisition. To address motion and stimulation artifacts with amplitudes reaching hundreds of mV while minimizing information loss, a sensor interface with high input range and fast artifacts recovery capability is essential.
View Article and Find Full Text PDFSensors (Basel)
March 2025
School of Science, Jiangnan University, Wuxi 214122, China.
In wireless sensor networks, sensors often collect and transmit a large amount of redundant data, which can lead to excessive battery consumption and subsequent performance degradation. To solve this problem, this paper proposes a Zoom-In Zoom-Out (ZIZO) method based on sparse vectors (SV-ZIZO). It operates in two parts: At the sensor level, given the temporal similarity of the data, a new compression method based on the sparse vector representation of segmented regions is proposed.
View Article and Find Full Text PDFAdv Physiol Educ
June 2025
Persist-AI Formulations Inc., West Sacramento, California, United States.
Dramatization, a teaching method where each student acts out or mimics a cell or body parts while the entire group represents the physiological process was adapted to produce original teaching videos paired with a pretest that activates memory and a posttest to prevent misconceptions. Three physiology instructors collaborated on Zoom to create six DramaZoom videos (Dramatization via Zoom) focused on hormone signaling with negative feedback in different contexts. In these videos, each instructor personalizes a different part of an organ system or a physiological process, which allows the visualization of complex concepts in endocrinology.
View Article and Find Full Text PDFCylindrical microlens arrays are important optical elements for autostereoscopic display. Conventional fixed focal length cylindrical microlens arrays do not allow switching between 2D mode and 3D mode when constructing a 3D viewing zone. In contrast, cylindrical liquid crystal microlens arrays with zoom characteristics allow switching between 2D and 3D states, as well as adjusting the width of the sub-viewing zone.
View Article and Find Full Text PDFCureus
December 2024
Psychiatry, Yale School of Medicine, New Haven, USA.
Introduction Following the COVID-19 pandemic, there was adoption of virtual psychotherapy. There are a number of benefits and drawbacks to telehealth video conferencing that are experienced by both clients and clinicians. The current qualitative study sought to outline the advantages and disadvantages that clients and clinicians have personally experienced in virtual versus in-person therapy in an effort to identify the reasons for which one medium may be preferred over another.
View Article and Find Full Text PDF