Article Abstract

Background: Arterial pulse wave velocity (PWV) is associated with increased mortality in aging and disease. Several studies have shown the accuracy of applanation tonometry carotid-femoral PWV (Cf-PWV) and the relevance of evaluating central aorta stiffness using 2D cardiovascular magnetic resonance (CMR) to estimate PWV, and aortic distensibility-derived PWV through the theoretical Bramwell-Hill model (BH-PWV). Our aim was to compare various methods of aortic PWV (aoPWV) estimation from 4D flow CMR, in terms of associations with age, Cf-PWV, BH-PWV and left ventricular (LV) mass-to-volume ratio while evaluating inter-observer reproducibility and robustness to temporal resolution.

Methods: We studied 47 healthy subjects (49.5 ± 18 years) who underwent Cf-PWV and CMR including aortic 4D flow CMR as well as 2D cine SSFP for BH-PWV and LV mass-to-volume ratio estimation. The aorta was semi-automatically segmented from 4D flow data, and mean velocity waveforms were estimated in 25 planes perpendicular to the aortic centerline. 4D flow CMR aoPWV was calculated: using velocity curves at two locations, namely ascending aorta (AAo) and distal descending aorta (DAo) aorta (S1, 2D-like strategy), or using all velocity curves along the entire aortic centreline (3D-like strategies) with iterative transit time (TT) estimates (S2) or a plane fitting of velocity curves systolic upslope (S3). For S1 and S2, TT was calculated using three approaches: cross-correlation (TTc), wavelets (TTw) and Fourier transforms (TTf). Intra-class correlation coefficients (ICC) and Bland-Altman biases (BA) were used to evaluate inter-observer reproducibility and effect of lower temporal resolution.

Results: 4D flow CMR aoPWV estimates were significantly (p < 0.05) correlated to the CMR-independent Cf-PWV, BH-PWV, age and LV mass-to-volume ratio, with the strongest correlations for the 3D-like strategy using wavelets TT (S2-TTw) (R = 0.62, 0.65, 0.77 and 0.52, respectively, all p < 0.001). S2-TTw was also highly reproducible (ICC = 0.99, BA = 0.09 m/s) and robust to lower temporal resolution (ICC = 0.97, BA = 0.15 m/s).

Conclusions: Reproducible 4D flow CMR aoPWV estimates can be obtained using full 3D aortic coverage. Such 4D flow CMR stiffness measures were significantly associated with Cf-PWV, BH-PWV, age and LV mass-to-volume ratio, with a slight superiority of the 3D strategy using wavelets transit time (S2-TTw).

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6907267PMC
http://dx.doi.org/10.1186/s12968-019-0584-xDOI Listing

Publication Analysis

Top Keywords

flow cmr
16
velocity curves
12
pulse wave
8
wave velocity
8
cardiovascular magnetic
8
magnetic resonance
8
mass-to-volume ratio
8
inter-observer reproducibility
8
cmr aopwv
8
aortic
6

Similar Publications

Fetal 4D Flow CMR for Advanced Diagnostics of Congenital Heart Disease: A Prospective Cohort Study.

Eur Heart J Cardiovasc Imaging

September 2025

Department of Diagnostic and Interventional Radiology, University Hospital Bonn, Venusberg-Campus 1, 53127, Bonn, Germany.

Aims: Fetal circulation undergoes complex changes in congenital heart disease (CHD) that are challenging to assess with fetal echocardiography. This study aimed to assess clinical feasibility and diagnostic value of 4D flow cardiac magnetic resonance (CMR) in fetal CHD.

Methods And Results: Pregnant women in advanced third trimester pregnancy with fetal CHD were prospectively recruited for fetal CMR between 08/2021 and 11/2024.

View Article and Find Full Text PDF

Full Free-Breathing Cardiac MRI: Enhancing Efficiency and Image Quality in Clinical Practice.

J Cardiovasc Magn Reson

September 2025

Department of Magnetic Resonance Imaging, Fuwai Hospital and National Center for Cardiovascular Diseases, Peking Union Medical College and Chinese Academy of Medical Sciences, Beijing 100037, China; Key Laboratory of Cardiovascular Imaging, Chinese Academy of Medical Sciences, Beijing 100730, China.

Background: Conventional cardiac magnetic resonance (CMR) examinations require patients to repeatedly hold their breath, which can reduce examination efficiency and pose challenges for patients unable to do so. This study aimed to demonstrate the feasibility and effectiveness of a full free-breathing CMR protocol in clinical practice.

Methods: Patients prospectively enrolled in this study underwent a full free-breathing CMR exam on a 3T scanner between June 1 and June 30, 2024.

View Article and Find Full Text PDF

PET/CT imaging of the late-gestation fetal brain in pregnant rats: A proof-of-concept study.

J Cereb Blood Flow Metab

September 2025

Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Harvard Medical School, Charlestown, MA, USA.

Preclinical PET studies offer the opportunity to elucidate molecular mechanisms underlying early neurodevelopment with minimal invasiveness. We demonstrated the feasibility of fetal brain PET in four pregnant rats ( = 42 fetuses). [F]FDG uptake in rat fetuses was readily visualized by PET imaging.

View Article and Find Full Text PDF

Aims: For many years, visual assessment has been the mainstay of detecting obstructive coronary artery disease (CAD) by stress perfusion cardiovascular magnetic resonance (S-CMR). Recently, fully automated quantitative assessment of myocardial blood flow (MBF) has been introduced. The value of MBF quantification in patients with coronary chronic total occlusion (CTO) is unknown.

View Article and Find Full Text PDF

Visually assessed ischaemia on cardiac magnetic resonance, but not quantitative perfusion metrics, predicts symptomatic improvement in coronary artery bypass.

J Cardiovasc Magn Reson

September 2025

Royal Brompton and Harefield Hospitals, part of Guy's and St Thomas' NHS Foundation Trust, London, UK; National Heart and Lung Institute, Imperial College London, UK. Electronic address:

Background: Serial perfusion cardiovascular magnetic resonance (CMR) in symptomatic patients undergoing coronary artery bypass grafting (CABG) may provide mechanistic insight into dynamic abnormalities of the myocardium.

Objectives: To assess how changes in cardiac reperfusion and remodelling associate with symptom improvement in patients undergoing CABG METHODS: Patients awaiting elective CABG completed serial quality of life questionnaires and detailed CMR at baseline and at 6-12 months post CABG as per protocol. Automated fully quantitative stress and rest myocardial blood flow was calculated, alongside assessment of the visual ischaemic burden.

View Article and Find Full Text PDF