Article Abstract

The nondestructive dosimetry achieved with electron paramagnetic resonance (EPR) dosimetry facilitates repetitive recording by the same dosimeter to increase the reliability of data. In precedent studies, solid paraffin was needed as a binder material to make the lithium formate monohydrate (LFM) EPR dosimeter stable and nonfragile; however, its use complicates dosimetry. This study proposes a newly designed pure LFM EPR dosimeter created by inserting LFM into a 3D-printed container. Dosimetric characteristics of the LFM EPR dosimeter and container, such as reproducibility, linearity, energy dependence, and angular dependence, were evaluated and verified through a radiation therapy planning system (RTPS). The LFM EPR dosimeters were irradiated using a clinical linear accelerator. The EPR spectra of the dosimeters were acquired using a Bruker EMX EPR spectrometer. Through this study, it was confirmed that there is no tendency in the EPR response of the container based on irradiation dose or radiation energy. The results show that the LFM EPR dosimeters have a highly sensitive dose response with good linearity. The energy dependence across each photon and electron energy range seems to be negligible. Based on these results, LFM powder in a 3D-printed container is a suitable option for dosimetry of radiotherapy. Furthermore, the LFM EPR dosimeter has considerable potential for in vivo dosimetry and small-field dosimetry via additional experiments, owing to its small effective volume and highly sensitive dose response compared with a conventional dosimeter.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s13246-019-00786-xDOI Listing

Publication Analysis

Top Keywords

lfm epr
24
epr dosimeter
16
3d-printed container
12
epr
10
lithium formate
8
formate monohydrate
8
electron paramagnetic
8
paramagnetic resonance
8
dosimetry radiotherapy
8
lfm
8

Similar Publications

Improvement of neutron sensitivity for lithium formate EPR dosemeters: a Monte Carlo analysis.

Radiat Prot Dosimetry

September 2023

Department of Physics and Chemistry 'Emilio Segrè', University of Palermo, Viale delle Scienze, Ed.18, I-90128 Palermo, Italy.

This work presents the computational analysis of the sensitivity improvements that could be achieved in lithium formate monohydrate (LFM) electron paramagnetic resonance (EPR) dosemeters exposed to neutron beams. Monte Carlo (MC) simulations were performed on LFM pellets exposed to neutron beams with different energy spectra at various depths inside a water phantom. Various computations were carried out by considering different enrichments of 6Li inside the LFM matrix as well as addition of different amounts of gadolinium oxide inside the pellet blend.

View Article and Find Full Text PDF

Suitability of superficial electron paramagnetic resonance dosimetry for in vivo measurement and verification of cumulative total doses during IMRT: A proof of principle.

Z Med Phys

November 2021

Klinik und Praxis fr Strahlentherapie am Klinikum Konstanz, Konstanz, Germany; Department of Radiation Oncology, Heidelberg University Hospital, Heidelberg, Germany; Clinical Cooperation Unit Molecular Radiation Oncology, German Cancer Research Center (DKFZ), Heidelberg, Germany.

Purpose: The present study investigates superficial in vivo dosimetry (IVD) by means of a previously proposed electron paramagnetic resonance (EPR) dosimetry system aiming at measuring and verifying total doses delivered by complex radiotherapy treatments. In view of novel regulatory requirements in Germany, differences between measured and planned total doses to the EPR dosimeters are analyzed and compared to reporting thresholds for significant occurrences.

Methods: EPR dosimeters, each consisting of one lithium formate monohydrate (LFM) and one polycrystalline l-alanine (ALA) pellet, were attached to the surface of an anthropomorphic head phantom.

View Article and Find Full Text PDF

In electron paramagnetic resonance (EPR) dosimetry, solid dosimeter materials such as alanine (AL) or, more recently, lithium formate monohydrate (LFM) are typically used. These materials offer high potential for applications in radiotherapy based on their favorable dosimetric properties. Nevertheless, EPR dosimetry is not widespread in the clinics.

View Article and Find Full Text PDF

Verification of lithium formate monohydrate in 3D-printed container for electron paramagnetic resonance dosimetry in radiotherapy.

Australas Phys Eng Sci Med

September 2019

Advanced Institute for Radiation Fusion Medical Technology, College of Medicine, The Catholic University of Korea, Annex B1 222 Banpo Road (Catholic University Seoul St. Mary's Hospital), Seocho-gu, Seoul, Korea.

The nondestructive dosimetry achieved with electron paramagnetic resonance (EPR) dosimetry facilitates repetitive recording by the same dosimeter to increase the reliability of data. In precedent studies, solid paraffin was needed as a binder material to make the lithium formate monohydrate (LFM) EPR dosimeter stable and nonfragile; however, its use complicates dosimetry. This study proposes a newly designed pure LFM EPR dosimeter created by inserting LFM into a 3D-printed container.

View Article and Find Full Text PDF