Article Abstract

Background Previously reported dual-energy CT methods for detecting noncalcified gallstones have reduced accuracy for gallstones smaller than 9 mm. Purpose To develop a dual-energy CT method for differentiating isoattenuating gallstones from bile and compare it with previously reported dual-energy CT methods by using a prospective ex vivo phantom reader study. Materials and Methods From May 2017 to May 2018, gallstones were collected from 105 patients (34 men; mean age, 51 years; age range, 18-84 years) undergoing cholecystectomy and placed inside 120-mL vials containing ox bile. The vials were placed inside a water-filled phantom and were scanned with dual-layer dual-energy CT. Thirty isoattenuating gallstones (4.3-24.7 mm in diameter) were evaluated. Conventional CT images, virtual noncontrast images, and monoenergetic images at 200 and 40 keV were created. Segmented images were created by using a two-dimensional histogram of Compton and photoelectric attenuation. Six readers evaluated the presence of isoattenuating gallstones in each image. Intra- and interreader agreement was measured by using percentage agreement, diagnostic performance was evaluated by using mean area under the receiver operating characteristic curve (AUC) estimates and pairwise comparisons, and the agreement of gallstone sizes measured at pathologic examination with those measured on segmented images was compared by using Bland-Altman analysis. Results For all gallstones, segmented images provided the highest mean intrareader (88.1%) and interreader (88.2% and 93.6%) agreements for all readers and reading sessions and the highest overall AUC (0.99; 95% confidence interval [CI]: 0.97, 1.00; adjusted < .02 for all). For gallstones larger than 9 mm, no significant difference was found between the segmented and monoenergetic AUCs (all > .94, adjusted > .05 for all). For gallstones measuring 9 mm or smaller, the segmented images had the highest overall AUC (0.99; 95% CI: 0.97, 1.00; adjusted < .01 for all). The mean difference in stone sizes was -0.6 mm, with limits of agreement from 2.6 to -3.8 mm. Conclusion Segmented images from Compton and photoelectric attenuation coefficients improve detection of isoattenuating gallstones compared with previously reported dual-energy CT methods. © RSNA, 2019 See also the editorial by Matos in this issue.

Download full-text PDF

Source
http://dx.doi.org/10.1148/radiol.2019190083DOI Listing

Publication Analysis

Top Keywords

isoattenuating gallstones
20
segmented images
20
reported dual-energy
12
dual-energy methods
12
gallstones
11
vivo phantom
8
images
8
compton photoelectric
8
photoelectric attenuation
8
highest auc
8

Similar Publications

Background: Because of its ability to identify isoattenuating gallstones, electron density (Rho) and effective atomic number (Z) maps derived from dual-energy computed tomography (CT) hold promise for secondary choledocholithiasis detection in lieu of magnetic resonance cholangiopancreatography (MRCP). The aim of this study was to investigate the diagnostic performance of dual-energy CT Rho/Z maps in detecting secondary choledocholithiasis in patients with gallbladder stones.

Materials And Methods: This retrospective single-center study included patients who underwent unenhanced abdominal dual-energy CT and MRCP for suspected secondary choledocholithiasis between July 2019 and March 2023.

View Article and Find Full Text PDF

Purpose: Impacted common bile duct stones cause severe acute cholangitis. However, the early and accurate diagnosis, especially iso-attenuating stone impaction, is still challenging. Therefore, we proposed and validated the bile duct penetrating duodenal wall sign (BPDS), which shows the common bile duct penetrating the duodenal wall on coronal reformatted computed tomography (CT), as a novel sign of stone impaction.

View Article and Find Full Text PDF

Background Previously reported dual-energy CT methods for detecting noncalcified gallstones have reduced accuracy for gallstones smaller than 9 mm. Purpose To develop a dual-energy CT method for differentiating isoattenuating gallstones from bile and compare it with previously reported dual-energy CT methods by using a prospective ex vivo phantom reader study. Materials and Methods From May 2017 to May 2018, gallstones were collected from 105 patients (34 men; mean age, 51 years; age range, 18-84 years) undergoing cholecystectomy and placed inside 120-mL vials containing ox bile.

View Article and Find Full Text PDF