Article Abstract

Background: Chest tube management protocols, particularly in patients with alveolar-pleural air leak due to recent surgery or trauma, are limited by concerns over safety, especially concerns about rapid and occult development of pneumothorax. A continuous, real-time monitor of pneumothorax could improve the quality and safety of chest tube management. We developed a rat model of pneumothorax to test a novel approach of measuring electrical impedance within the pleural space as a monitor of lung expansion.

Materials And Methods: Anesthetized Sprague-Dawley rats underwent right thoracotomy. A novel impedance sensor and a thoracostomy tube were introduced into the right pleural space. Pneumothorax of varying volumes ranging from 0.2 to 20 mL was created by syringe injection of air via the thoracostomy tube. Electrical resistance measurements from the pleural sensor and fluoroscopic images were obtained at baseline and after the creation of pneumothorax and results compared.

Results: A statistically significant, dose-dependent increase in electrical resistance was observed with increasing volume of pneumothorax. Resistance measurement allowed for continuous, real-time monitoring of pneumothorax development and the ability to track pneumothorax resolution by aspiration of air via the thoracostomy tube. Pleural resistance measurement demonstrated 100% sensitivity and specificity for all volumes of pneumothorax tested and was significantly more sensitive for pneumothorax detection than fluoroscopy.

Conclusions: The electrical impedance-based pleural space sensor described in this study provided sensitive and specific pneumothorax detection, which was superior to radiographic analysis. Real-time, continuous monitoring for pneumothorax has the potential to improve the safety, quality, and efficiency of postoperative chest tube management.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jss.2018.05.005DOI Listing

Publication Analysis

Top Keywords

pneumothorax
13
chest tube
12
tube management
12
pleural space
12
thoracostomy tube
12
electrical impedance
8
continuous real-time
8
air thoracostomy
8
electrical resistance
8
resistance measurement
8

Similar Publications

Background: Percutaneous transthoracic lung biopsy (PTNB) guided by Computed Tomography (CT) greatly depends on the operators' skill for accuracy. This study aimed to evaluate whether three-dimensionally(3D) printed navigational templates for percutaneous transthoracic lung biopsy achieve diagnostic yield comparable to conventional computed tomography guidance.

Materials And Methods: Conducted from 1 November 2020, to 27 July 2023, this noninferiority randomized clinical trial included 159 patients with peripheral lung masses (≥30 mm).

View Article and Find Full Text PDF

Background: Acupuncture is a widely recognized complementary therapy with proven therapeutic benefits; however, concerns regarding patient safety persist due to adverse events ranging from minor complications to severe outcomes like pneumothorax and nerve injury.

Objective: This study aims to identify common adverse events in acupuncture, propose innovative risk mitigation strategies, establish standardized best practices, enhance practitioner training, and examine global disparities in safety protocols.

Design: Comprehensive review of existing literature, clinical case studies, and international safety guidelines.

View Article and Find Full Text PDF

Conservative Management of Non-Tension Pneumothorax Following Drainage of Pleural Empyema.

Pediatr Pulmonol

September 2025

Department of Paediatric Respiratory Medicine, Staffordshire Children's Hospital at Royal Stoke, University Hospitals of North Midlands NHS Trust, Stoke on Trent, UK.

Pleural empyema is a recognized complication of pneumonia and causes significant morbidity in children. Insertion of a small-bore chest drain shortens hospital admission but can be associated with pneumothorax. This is usually assumed to be caused by a bronchopleural fistula or a displaced drain and therefore under pressure, requiring surgical intervention.

View Article and Find Full Text PDF

Objective: Persistent pulmonary air leak happens in a minority of patients with various thoracic pathologies. Spiration (Olympus America Inc.) bronchial valves (BVs) are approved by the Food and Drug Administration under Humanitarian Use Device status to treat persistent air leak after lung resection.

View Article and Find Full Text PDF