98%
921
2 minutes
20
Objective: Endovascular aneurysm sealing (EVAS) with the Nellix stent graft system is a novel concept in the management of abdominal aortic aneurysm (AAA) that aims to reduce the prevalence of all endoleaks following endovascular repair. There are few data describing the longer-term durability of this approach. The aim was to report the longer-term outcomes following EVAS in a single centre.
Methods: This is a retrospective review of all patients that underwent Nellix at Cambridge University Hospitals Foundation Trust. Factors that are described as device failure include secondary sac rupture, graft explantation, further surgical procedures for Type 1 endoleak, or major migration of the stent grafts with pressurisation of the aortic sac.
Results: A total of 161 patients have been treated with Nellix. The indications included primary AAA (n = 115), ruptured AAA (n = 4), salvage of other aortic grafts (n = 18), primary iliac aneurysm (n = 6), and chimney EVAS (ChEVAS) for pararenal AAA (n = 18). In total there have been 42 graft failures in patients treated with EVAS for primary AAA. The 4 year freedom from graft failure was 42% in patients treated for primary AAA. Failures mostly occurred more than 2 years post-Nellix implant. There were eight secondary sac ruptures (incidence 2.4 per 100 person years) and there have been 14 graft explants.
Conclusions: Failure of aneurysm sealing following treatment with Nellix has been more common than anticipated and can cause aortic rupture. Post-operative surveillance of Nellix stent grafts is crucial to identify features of failure.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.ejvs.2018.06.031 | DOI Listing |
J Clin Med
August 2025
Division of Vascular Surgery, 2nd Department of Surgery, Faculty of Medicine, "G. Gennimatas" General Hospital of Thessaloniki, School of Health Sciences, Aristotle University of Thessaloniki, 54124 Thessaloniki, Greece.
The aim of our study was to document and analyze the long-term geometric alterations that occur in the infrarenal aorta and iliac arteries over time after a successful elective standard endovascular abdominal aneurysm repair (EVAR) as well as to investigate the potential relationship of aortoiliac remodeling with the long-term complications of EVAR. The prospectively collected clinical and computed tomography angiography (CTA) data from 168 patients treated with elective standard EVAR between 2013 and 2018 were retrospectively analyzed. Follow-up assessments were performed at 1, 24, and 60 months postoperatively.
View Article and Find Full Text PDFEur J Vasc Endovasc Surg
August 2025
Division of Vascular and Endovascular Surgery, Department of Cardiac, Thoracic, Vascular Sciences and Public Health, University of Padua, Padua, Italy.
Objective: This study aimed to report the 2 year outcomes obtained with the E-nside endograft (Artivion) for the treatment of complex abdominal aortic aneurysm (cAAA) and thoraco-abdominal aortic aneurysm (TAAA).
Methods: This was a prospective, multicentre study. Data from the multicentre ItaliaN Branched Registry of E-nside EnDograft (INBREED) were analysed.
J Clin Med
July 2025
Department of Vascular Surgery, Rigshospitalet, Blegdamsvej 9, 2100 Copenhagen, Denmark.
: Fenestrated stentgrafting has become a first-line treatment for juxtarenal aneurysms, and the incorporation of all renovisceral vessels with fenestrations has become common to increase the proximal sealing zone. This increases the complexity of the repair compared to using fewer fenestrations, and stenting of the celiac artery (CA), in particular, can be technically challenging. : This study evaluates the mid-term outcomes of leaving the celiac artery unstented during quadruple fenestrated stentgrafting for complex aortic aneurysms.
View Article and Find Full Text PDFEJVES Vasc Forum
May 2025
Department of Vascular Surgery, University Medical Centre Utrecht, Utrecht, the Netherlands.
EJVES Vasc Forum
May 2025
Department of Vascular Surgery, Innlandet Hospital Trust, Hamar, Norway.
Objective: The Nellix endovascular aneurysm sealing (EVAS) system demonstrated promising short and midterm outcomes following its introduction in 2011. However, long term results have shown a high incidence of graft failure. This study aimed to compare the long term outcomes of EVAS compared with standard endovascular aneurysm repair (EVAR).