Article Abstract

Introduction: A comparison of precision of intraocular implant power calculation by computing keratometry values by two different methods, namely, the automated and manual keratometry (MK), was done. For checking this accuracy, the parameter taken into consideration was the absolute refractive error which was ascertained postoperatively.

Subjects And Methods: This study was conducted in the Department of Ophthalmology, Sri Aurobindo Institute of Medical Sciences and PG Institute, Indore. At a tertiary eye care centre in Central India (Sri Aurobindo Institute of Medical Sciences, Indore). Duration of the study was 18 months. Sample size was 66 individuals who were with cataract (nuclear sclerosis I to III). Although keratometry was done by both methods and implant power derived separately by computing both readings, decision of which power to be implanted in an eye would depend on the group, in which patient would fall. Group A were prospective candidates who would be implanted intraocular lens (IOL) of that power as assessed by computing MK value, whereas Group B were participants who would be implanted IOL of that power as assessed by computing value obtained by automated keratometry (AK). First patient fell in Group A and second in Group B, third again in Group A till 33 patients had been operated in each group.

Results: Bland-Altman plot thus obtained showed that the two keratometers are comparable. The postoperative refractive errors for the two Groups (A and B), showed that if an error of ±0.50 D or less is considered, then in Group A 81% of patients achieved this and 87% of patients required this as spectacle aid in Group B.

Conclusions: In this study, we compared the accuracy of AK with that of MK for calculation of implant power. It was concluded from this study that AK is a simple keratometric technique that appeared to be more accurate than MK.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6055314PMC
http://dx.doi.org/10.4103/tjo.tjo_58_17DOI Listing

Publication Analysis

Top Keywords

implant power
12
automated manual
8
manual keratometry
8
intraocular lens
8
power calculation
8
sri aurobindo
8
aurobindo institute
8
institute medical
8
medical sciences
8
group
8

Similar Publications

Rationale: Radial artery spasm (RAS) is a common complication during transradial cerebral angiography (TRA), but currently, the optimal prevention strategy is not well established. Papaverine has anti-vasospasm, sedative and analgesic effects. However, the efficacy of papaverine in preventing RAS during TRA remains unknown.

View Article and Find Full Text PDF

Dopamine (DA) signaling is essential for neurodevelopment and is particularly sensitive to disruption during adolescence. Protein restriction (PR) can impair DA dynamics, yet mechanistic insights remain limited due to challenges in real-time neurochemical sensing. Here, we present aptCFE, a robust implantable aptamer-based sensor fabricated via a reagent-free, 3 min electrochemical conjugation (E-conjugation) using amine-quinone chemistry.

View Article and Find Full Text PDF

Topic: To compare the outcomes of surgical approaches to correct ametropia following cataract and lens surgery.

Clinical Relevance: Despite advancements in the field of biometry and intraocular lens (IOL) power calculation formulas, complete elimination of refractive surprises following cataract and lens surgery is impossible. Preferred Practice Patterns acknowledges the possibility of refractive surprise following cataract surgery; however, no recommendations regarding the preferred treatment have been given.

View Article and Find Full Text PDF

Transient electronics that can degrade after fulfilling their designed functionalities offer transformative potentials in biomedical implants (eliminating secondary surgeries), ecofriendly consumer electronics (reducing e-waste), and secure systems. However, the development of reliable transient energy supplies remains a critical challenge, thus limiting their widespread implementation. Among various solutions, wireless power supplies via near-field inductive coupling stand out as particularly promising candidates.

View Article and Find Full Text PDF

We report the observation of negative differential resistance (NDR) in single-atom single-electron devices based on arsenic, phosphorus and potassium dopants implanted in a silicon host matrix. All devices exhibit NDR, with the potassium-based one exhibiting NDR at room temperature because of the larger charging and confinement energies. Our experimental results are reproduced with a simple model that assumes sequential electron tunnelling through two series-connected charge centres, each having two discrete energy levels.

View Article and Find Full Text PDF