Independent dose validation system for Gamma Knife radiosurgery, using a DICOM-RT interface and Geant4.

Phys Med

Department of Radiation Convergence Engineering, Yonsei University, Wonju, South Korea. Electronic address:

Published: July 2018


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Leksell GammaPlan was specifically designed for Gamma Knife (GK) radiosurgery planning, but it has limited accuracy for estimating the dose distribution in inhomogeneous areas, such as the embolization of arteriovenous malformations. We aimed to develop an independent patient dose validation system based on a patient-specific model, constructed using a DICOM-RT interface and the Geant4 toolkit. Leksell Gamma Knife Perfexion was designed in Geant4.10.00 and includes a DICOM-RT interface. Output factors for each collimator in a sector and dose distributions in a spherical water phantom calculated using a Monte Carlo (MC) algorithm were compared with the output factors calculated by the tissue maximum ratio (TMR) 10 algorithm and dose distributions measured using film, respectively. Studies using two types of water phantom and two patient simulation cases were evaluated by comparing the dose distributions calculated by the MC, the TMR and the convolution algorithms. The water phantom studies showed that if the beam size is small and the target is located in heterogeneous media, the dose difference could be up to 11%. In the two patient simulations, the TMR algorithm overestimated the dose by about 4% of the maximum dose if a complex and large bony structure was located on the beam path, whereas the convolution algorithm showed similar results to those of the MC algorithm. This study demonstrated that the in-house system could accurately verify the patient dose based on full MC simulation and so would be useful for patient cases where the dose differences are suspected.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejmp.2018.06.008DOI Listing

Publication Analysis

Top Keywords

gamma knife
12
dicom-rt interface
12
dose distributions
12
water phantom
12
dose
10
dose validation
8
validation system
8
knife radiosurgery
8
interface geant4
8
patient dose
8

Similar Publications

Introduction Stereotactic radiosurgery (SRS) is widely regarded as the standard of care after the resection of brain metastases in order to reduce local cavity recurrence risk. The objective of this study was to explore the reproducibility of published outcomes for patients receiving post-operative stereotactic radiosurgery (cavity SRS) in a National Health Service (NHS) setting for a non-selective series of patients. For our service, the median interval between surgery to cavity SRS (cSRS) is eight weeks, whereas similar timelines have been found to have a deleterious impact on survival in the published literature.

View Article and Find Full Text PDF

Background: Stereotactic Body Radiotherapy (SBRT) has been proven to be a safe and effective alternative to surgery in patients with metastatic primary sarcoma. However, data describing tumor response in relation to the given radiotherapy dose is lacking. Therefore, this study aims at analyzing efficacy and dose-response relationship in a retrospective cohort.

View Article and Find Full Text PDF

Introduction: Radiosurgery targeting the thalamus has long been used to treat refractory pain, with medial thalamotomy as a key approach. Traditionally, targeting relied on indirect methods based on anatomical atlases, which do not account for individual variations in brain connectivity. Recent advances in connectomic-guided stereotactic radiosurgery have improved precision in the treatment of movement disorders, but their application to pain management remains underexplored.

View Article and Find Full Text PDF

While World Health Organization (WHO) grade I meningiomas are typically slow growing and associated with favorable prognoses, a subset may exhibit unexpectedly aggressive behavior and resistance to conventional treatment approaches. Recurrent grade I meningiomas, in particular, are associated with a poorer prognosis despite their benign histological classification, underscoring the need for advanced genomic and radiomic analyses to refine diagnostic accuracy. We present a case of a 52-year-old female with a grade I parafalcine meningioma initially deemed nonaggressive, but ultimately recurred multiple times over several years despite undergoing repeated craniotomies and several courses of radiosurgery.

View Article and Find Full Text PDF

Management of recurrent adenoid cystic carcinoma (ACC) in elderly patients remains challenging due to comorbidities, functional impairments, and anatomically complex tumor locations that complicate surgical access and increase operative risk. The ZAP-X Gyroscopic Radiosurgery System (ZAP Surgical Systems, Inc., San Carlos, CA, USA) offers a highly precise, non-invasive treatment modality, potentially suitable for salvage therapy in previously irradiated fields and in medically inoperable patients.

View Article and Find Full Text PDF