Article Abstract

Purpose: The 11-gauge (11G) stereotactic vacuum-assisted breast biopsy (VABB) showed a better profile than 14G-VABB in terms of feasibility, safety, microcalcification sampling, and accuracy. Underestimation rates were significantly lower with 11G-VABB than with 14G-VABB. Thus, the introduction of an even larger needle at the VABB procedure could reduce this rate further. The purpose of this study was to compare the overall performance of stereotactic VABB with 8G and 11G needles.

Materials And Methods: Four hundred and three VABBs performed between July 2012 and February 2015 at the Breast Diagnostic Unit of Careggi Hospital in Florence were retrospectively analyzed; 197 were performed with 11G-VABB and 206 with 8G-VABB. Lesions were classified according to mammographical patterns in microcalcifications, architectural distortions, or opacities, and all biopsy targets were classified according to BIRADS classification as BIRADS III, IV or V. Data were collected on radiological classification of targets, imaging presentation, procedure time, number of specimens per procedure, and microcalcification retrieval on histological findings. Surgery was always performed when high-risk or malignant lesions (B3 or B5) were detected; the final diagnosis was made on surgical pathology.

Results: Compared to VABB with an 11G needle, 8G-VABB allows a reduction in the time needed to complete the procedure (20.6 versus 27.4, P < 0.00001) and the number of specimens collected per lesion (21.6 versus 12.2, P < 0.00001). Moreover, 8G-VABB resulted in the same diagnostic accuracy, and the underestimation rates were comparable between the two groups for both B3 and DCIS lesions.

Conclusions: The 8G needle should be considered as a valid alternative option in VABB for breast lesions.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejso.2017.07.011DOI Listing

Publication Analysis

Top Keywords

stereotactic vacuum-assisted
8
vacuum-assisted breast
8
breast biopsy
8
vabb 11g
8
biopsy comparison
4
comparison 11-
4
11- 8-gauge
4
8-gauge needles
4
needles purpose
4
purpose 11-gauge
4

Similar Publications

Clinical implications of clip migration after stereotactic-assisted vacuum-assisted breast biopsies.

Eur J Surg Oncol

August 2025

Breast Care Unit, Department of Obstetrics and Gynecology, University of Rostock, Germany; Department of Diagnostic and Interventional Radiology, Klinikum Suedstadt Rostock, Germany.

A stereotactic vacuum-assisted breast biopsy (VABB) is used to assess non-palpable breast lesions. The current study aims to identify possible influencing factors for clip migration and the impact of clip migration on surgery. In a retrospective study, clip migration in post-biopsy mammograms was evaluated (using a cut-off of ≥10 mm vs.

View Article and Find Full Text PDF

Aim: Stereotactic-guided vacuum-assisted biopsy (STVAB) is the standard of care to biopsy mammographically detected breast lesions without a sonographic correlate. It can be performed using the conventional (CBA) or lateral-arm biopsy approach (LABA). Tissue marker (clip) placement is routinely done post-biopsy.

View Article and Find Full Text PDF

Background: Compared with traditional mammography, digital breast tomosynthesis (DBT) increases the detection of breast lesions by reducing the overlapping masking effect of glands through multi-angle exposures of a rotating X-ray tube. Breast biopsy under the guidance of mammography can effectively diagnose breast lesions. The main methods of mammography-guided biopsy include wire positioning, core needle biopsy (CNB), and vacuum-assisted breast biopsy (VABB).

View Article and Find Full Text PDF

Mammography is an essential tool in breast screening, often revealing lesions that appear as microcalcifications with or without an associated mass. Decisions about biopsy requirements are guided by the BI-RADS system, aiming to confirm the histopathology of suspicious lesions while avoiding unnecessary procedures. A vacuum-assisted breast biopsy (VABB) is a minimally invasive procedure for diagnosing breast abnormalities.

View Article and Find Full Text PDF

Architectural distortions (ADs) detected on digital breast tomosynthesis (DBT) are more frequently associated with nonmalignant pathologies than those detected on digital mammography. The purpose of this study was to evaluate outcomes of ADs detected by DBT alone with no ultrasound correlate and nonmalignant results on core needle biopsy (CNB) when adopting a management strategy selectively incorporating imaging surveillance. This retrospective study included patients with ADs detected by DBT alone with no ultrasound correlate and nonmalignant results on stereotactic CNB with 12 vacuum-assisted 9-gauge cores from July 1, 2020, to December 31, 2023.

View Article and Find Full Text PDF