A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 197

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3165
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 317
Function: require_once

Evaluation of the HeartWare ventricular assist device Lavare cycle in a particle image velocimetry model and in clinical practice. | LitMetric

Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Objectives: Ventricular blood stasis is a concern for continuous flow mechanical support devices and might contribute to the formation of thromboembolic events. The HeartWare® Ventricular Assist System (HVAD®) is equipped with the Lavare™ cycle that is a periodic speed modulation feature designed to alter flow patterns within the left ventricle and reduce areas of potential blood stasis. Here, we report in vitro and clinical findings on the effects of the Lavare cycle.

Methods: The effect of pump speed changes on the intraventricular flow field was examined with an in vitro particle image velocimetry model. The clinical impact of the Lavare cycle was evaluated through a retrospective review of the ReVOLVE study which includes 248 patients implanted with the HVAD following Conformité Européenne Mark in nine centres in Europe and Australia. Baseline characteristics, adverse event profiles and Kaplan-Meier survival estimates were stratified by patients using/not using the Lavare cycle.

Results: Particle image velocimetry showed increased ventricular washout with an active Lavare cycle as measured by the fluid velocities and angular dispersion parameters. With the Lavare cycle on, there was also a 22% decrease in the stagnation index compared with when the Lavare cycle was off. In the ReVOLVE registry, patients with the Lavare cycle turned on (n = 215) were supported for 497 patient-years, whereas patients who did not use the speed modulation (n = 33) were supported for 39.3 patient-years. The Lavare cycle did not significantly affect patient survival as both groups had approximately an 80% survival after 1 year. Patients using the Lavare cycle had significantly fewer rates of stroke [0.06 vs 0.20 events per patient-year (EPPY), P = 0.0008], sepsis (0.03 vs 0.15 EPPY, P = 0.0003) and right heart failure (0.03 vs 0.18 EPPY, P < 0.0001) with no difference in the transplant or recovery rates among the two cohorts.

Conclusions: The Lavare cycle effectively generates ventricular washout and the adverse event profiles of ReVOLVE patients with the Lavare cycle on were better than those with the Lavare cycle off. Larger studies are warranted to verify the positive effect of the Lavare cycle and to optimize speed modulation settings, so additional clinically relevant improvements can be realized.

Download full-text PDF

Source
http://dx.doi.org/10.1093/ejcts/ezw232DOI Listing