Article Abstract

Purpose: Common camera calibration methods employed in current laparoscopic augmented reality systems require the acquisition of multiple images of an entire checkerboard pattern from various poses. This lengthy procedure prevents performing laparoscope calibration in the operating room (OR). The purpose of this work was to develop a fast calibration method for electromagnetically (EM) tracked laparoscopes, such that the calibration can be performed in the OR on demand.

Methods: We designed a mechanical tracking mount to uniquely and snugly position an EM sensor to an appropriate location on a conventional laparoscope. A tool named fCalib was developed to calibrate intrinsic camera parameters, distortion coefficients, and extrinsic parameters (transformation between the scope lens coordinate system and the EM sensor coordinate system) using a single image that shows an arbitrary portion of a special target pattern. For quick evaluation of calibration results in the OR, we integrated a tube phantom with fCalib prototype and overlaid a virtual representation of the tube on the live video scene.

Results: We compared spatial target registration error between the common OpenCV method and the fCalib method in a laboratory setting. In addition, we compared the calibration re-projection error between the EM tracking-based fCalib and the optical tracking-based fCalib in a clinical setting. Our results suggest that the proposed method is comparable to the OpenCV method. However, changing the environment, e.g., inserting or removing surgical tools, might affect re-projection accuracy for the EM tracking-based approach. Computational time of the fCalib method averaged 14.0 s (range 3.5 s-22.7 s).

Conclusions: We developed and validated a prototype for fast calibration and evaluation of EM tracked conventional (forward viewing) laparoscopes. The calibration method achieved acceptable accuracy and was relatively fast and easy to be performed in the OR on demand.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5290310PMC
http://dx.doi.org/10.1007/s11548-016-1406-3DOI Listing

Publication Analysis

Top Keywords

calibration evaluation
8
electromagnetically tracked
8
augmented reality
8
calibration
8
fast calibration
8
calibration method
8
laparoscopes calibration
8
coordinate system
8
opencv method
8
fcalib method
8

Similar Publications

ISO 21043 is a new international standard for forensic science. It provides requirements and recommendations designed to ensure the quality of the forensic process. It includes Parts on: 1 vocabulary; 2 recovery, transport, and storage of items; 3 analysis; 4 interpretation; and 5 reporting.

View Article and Find Full Text PDF

Assessing the importance of functional form selection in developing calibration functions for the highway safety manual predictive models.

J Safety Res

September 2025

School of Civil Engineering and Environmental Science, University of Oklahoma, Norman, OK 73019, United States; School of Industrial and Systems Engineering, University of Oklahoma, Norman, OK 73019, United States. Electronic address:

Introduction: The Highway Safety Manual (HSM) offers crash prediction models for estimating the number and severity of crashes for several facility types. However, since these models were developed using data from selected states within the United States, they should be calibrated when applied to a new jurisdiction. The HSM recommended using a scalar calibration factor to calibrate the prediction models.

View Article and Find Full Text PDF

Accurate determination of the parameters of each high purity germanium, HPGe detectors ensure the precision of quantitative results obtained from spectrum analysis. This study presents a comprehensive performance evaluation and long-term quality control assessment of a high-purity germanium (HPGe) gamma spectrometry system that has been operational for over 15 years. Key spectrometric measures were recorded, including energy resolution, peak shape ratios, asymmetry, peak-to-Compton ratio, relative efficiency, electronic noise, minimum detectable activity (MDA), and repeatability and reproducibility of the system.

View Article and Find Full Text PDF

Coronary heart disease (CHD) is a leading cause of morbidity and mortality; patients with type 2 diabetes mellitus (T2DM) are at particularly high risk, highlighting the need for reliable biomarkers for early detection and risk stratification. We investigated whether combining the stress hyperglycemia ratio (SHR) and systemic inflammation response index (SIRI) improves CHD detection in T2DM. In this retrospective cohort of 943 T2DM patients undergoing coronary angiography, associations of SHR and SIRI with CHD were evaluated using multivariable logistic regression and restricted cubic splines; robustness was examined with subgroup and sensitivity analyses.

View Article and Find Full Text PDF

Population-wide screening may accelerate the decline of tuberculosis (TB) incidence, but the optimal screening algorithm and duration must weigh resource considerations. We calibrated a deterministic transmission model to TB epidemiology in Viet Nam. We simulated three population-wide screening algorithms from 2025: sputum nucleic acid amplification tests (NAAT, Xpert MTB/RIF Ultra) only; chest radiography (CXR) followed by NAAT; and CXR-only without microbiological confirmation.

View Article and Find Full Text PDF