Severity: Warning
Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 197
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1075
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3195
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 317
Function: require_once
98%
921
2 minutes
20
New left ventricular assist devices (LVADs) offer both important advantages and potential hazards. VAD development requires better and expeditious ways to identify these advantages and hazards. We validated in an isolated working heart the hemodynamic performance of an intraventricular LVAD and investigated how its outflow cannula interacted with the aortic valve. Hearts from six pigs were explanted and connected to an isolated working heart setup. A miniaturized LVAD was implanted within the left ventricle (tMVAD, HeartWare Inc., Miami Lakes, FL, USA). In four experiments blood was used to investigate hemodynamics under various loading conditions. In two experiments crystalloid perfusate was used, allowing visualization of the outflow cannula within the aortic valve. In all hearts the transapical miniaturized ventricular assist device (tMVAD) implantation was successful. In the blood experiments hemodynamics similar to those observed clinically were achieved. Pump speeds ranged from 9 to 22 krpm with a maximum of 7.6 L/min against a pressure difference between ventricle and aorta of ∼50 mm Hg. With crystalloid perfusate, central positioning of the outflow cannula in the aortic root was observed during full and partial support. With decreasing aortic pressures the cannula tended to drift toward the aortic root wall. The tMVAD could unload the ventricle similarly to LVADs under conventional cannulation. Aortic pressure influenced central positioning of the outflow cannula in the aortic root. The isolated heart is a simple, accessible evaluation platform unaffected by complex reactions within a whole, living animal. This platform allowed detection and visualization of potential hazards.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/aor.12730 | DOI Listing |