Purpose: This study aimed to investigate the feasibility of efficiently using a rigid image registration (RIR) algorithm or a deformable image registration (DIR) algorithm to match medical images and evaluate the impact of setup errors on intensity modulated radiation therapy of lung cancer patients.
Methods: Ten lung cancer patients were chosen randomly each day and were subjected to image-guided radiotherapy. The clinical registration between cone-beam computed tomography (CBCT) images and treatment planning system CT images was performed by applying both RIR and DIR; the clinical registration was evaluated on the basis of the contour index, including dice similarity coefficient, sensitivity, and positive predictive value; the optimal scheme of image registration was selected to ensure that the actual irradiation isocenter was consistent with the treatment planning isocenter. In each patient, the translational errors in the right-left (x), superior-inferior (y), and anterior-posterior (z) directions and the rotational errors in the u, υ, and w directions formed by the x, y, and z directions were calculated and analyzed daily in the whole course of treatment; margins were calculated according to this equation: M = 2.5∑+ 0.7δ.
Results: The tumors and the surrounding soft tissues of the patients are shown more clearly in the CBCT images than in the CT images. DIR can be applied more efficiently than RIR to determine the morphological and positional changes in the organs shown in the images with the same or different modalities in the different period. The setup errors in translation in the x, y and z axes were 0.05±0.16, 0.09±0.32 and -0.02±0.13 cm, respectively; by contrast, the setup errors in rotation in u, υ and w directions were (0.41±0.64)°, (-0.08±0.57)° and (-0.03±0.62)°, respectively. The setup errors in the x, y and z axes of the patients indicated that the margins expansions were 0.82, 1.15 and 0.72 cm, respectively.
Conclusion: CBCT with DIR can measure and correct the setup errors online; as a result, setup errors in lung cancer treatments can be significantly reduced and the accuracy of radiotherapy can be enhanced.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.3233/XST-160568 | DOI Listing |
Phys Med
September 2025
Department of Biomedical Technology, Faculty of Medicine and Health Technology, Tampere University, Korkeakoulunkatu 8, 33720 Tampere, Finland; School of Electrical and Information Engineering, University of the Witwatersrand, Johannesburg, Jan Smutslaan 1, 2050 Braamfontein, South Africa.
Background And Objective: Correction of rotational setup errors by tilting the treatment couch improves target dose accuracy and prevents healthy tissue overdosage in external beam radiotherapy. However, couch tilts may cause secondary patient motion. This study aimed to quantify the secondary motion caused by pitch and roll corrections and to evaluate the feasibility of surface imaging for detecting the secondary motion in pelvic radiotherapy.
View Article and Find Full Text PDFMed Phys
September 2025
Department of Accelerator and Medical Physics, National Institutes for Quantum Science and Technology (QST), Chiba, Japan.
Background: Multi-ion radiotherapy using carbon, oxygen, and neon ions aims to improve local control by increasing dose-averaged linear energy transfer (LET) in the target. However, there has been limited understanding of how to utilize variables for multi-ion treatment planning such as the selection and arrangement of ion species.
Purpose: An in silico study was conducted to explore the feasibility of increasing a minimum LET, and the optimal selection and arrangement of ion species in multi-ion therapy for increasing LET in tumors of varying sizes mimicking bone and soft tissue sarcomas (BSTS).
Anal Chim Acta
November 2025
Institute of Nano Science and Technology, Knowledge City, Sahibzada Ajit Singh Nagar, Sector- 81, Punjab, 140306, India. Electronic address:
Background: Iron (Fe) is an essential micronutrient for plant growth, but the conventional DTPA soil analysis method for detecting available iron has notable limitations, requiring advanced instruments and lengthy preparation time. Developing a more affordable, user-friendly, and efficient method for iron detection in soil could greatly improve crop nutrition management. Here, a facile nanoscopic method was developed to quantify available Fe ions in the soil by forming a luminescence quenching complex in chelation with bathophenanthroline disulphonic acid disodium salt (Fe/BPDS complex).
View Article and Find Full Text PDFInt J Comput Assist Radiol Surg
September 2025
Department of Rhythmology, University Heart Center Lübeck, University Hospital Schleswig-Holstein, Ratzeburger Allee 160, Lübeck, 23652, Germany.
Purpose: Ultrasound (US) is commonly used to assess left ventricular motion for examination of heart function. In stereotactic arrhythmia radioablation (STAR) therapy, managing cardiorespiratory motion during radiation delivery requires representation of motion information in computed tomography (CT) coordinates. Similar to conventional US-guided navigation during surgical procedures, 3D US can provide real-time motion data of the radiation target that could be transferred to CT coordinates and then be accounted for by the radiation system.
View Article and Find Full Text PDFTech Innov Patient Support Radiat Oncol
September 2025
Princess Margaret Cancer Centre, University Health Network, Toronto, ON, Canada.
Background: Tattoos help guide field placement in breast re-irradiation. This study evaluates the stability of medial tattoos in patients with prior breast radiotherapy (RT) to determine their reliability as surface markers.
Materials And Methods: We retrospectively identified patients who had breast/chest wall re-irradiation between January 2022 and December 2023 (RT) and prior breast RT (RT) at our institution.