Article Abstract

A total integrated genetic analysis microsystem was developed, which consisted of solid phase extraction (SPE), polymerase chain reaction (PCR), and immunochromatographic strip (ICS) parts for multiplex colorimetric detection of pathogenic Staphylococcus aureus (S. aureus) and Escherichia coli O157:H7 (E. coli O157:H7) on a portable genetic analyzer. Utilizing a slidable chamber, which is a movable glass wafer, complex microvalves could be eliminated for fluidic control in the microchannel, which could simplify the chip design and chip operation. The integrated slidable microdevice was composed of 4 layers: a 4-point Pt/Ti resistance temperature detector (RTD) wafer, a micro-patterned channel wafer, a 2 μL volume slidable chamber, and an ICS. The entire process from the DNA extraction in the SPE chamber to the detection of the target gene expression by the ICS was serially performed by simply sliding the slidable chamber from one part to another functional part. The total process for multiplex pathogenic S. aureus and E. coli O157:H7 detection on the integrated slidable microdevice was accomplished within 55 min with a detection limit of 5 cells. Furthermore, spiked bacteria samples in milk were also successfully analysed on the portable genetic analysis microsystem with sample-in-answer-out capability. The proposed total integrated microsystem is adequate for point-of-care DNA testing in that no microvalves and complex tubing systems are required due to the use of the slidable chamber and the bulky and expensive fluorescence or electrochemical detectors are not necessary due to the ICS based colorimetric detection.

Download full-text PDF

Source
http://dx.doi.org/10.1039/c5lc00801hDOI Listing

Publication Analysis

Top Keywords

slidable chamber
16
integrated slidable
12
coli o157h7
12
solid phase
8
phase extraction
8
polymerase chain
8
chain reaction
8
immunochromatographic strip
8
parts multiplex
8
multiplex colorimetric
8

Similar Publications

A novel low-cost and simple fabrication technique for a paper-based analytical device using super glue.

Anal Chim Acta

November 2024

Major of Biomedical Engineering, Division of Smart Healthcare, College of Information Technology and Convergence, Pukyong National University, Republic of Korea; Industry 4.0 Convergence Bionics Engineering, Pukyong National University, Busan, 48513, Republic of Korea. Electronic address: jhshin@pkn

Article Synopsis
  • Microfluidic paper-based analytical devices (μPADs) are praised for being affordable and portable tools for point-of-care testing, but traditional fabrication methods can be complex and costly, particularly in low-resource environments.
  • This research presents a new fabrication method for μPADs using 3D-printed chambers and super glue vapor, allowing for quick and easy production, effective in just 5 minutes without specialized training.
  • The study's significance lies in its introduction of a simplified way to create μPADs for detecting various analytes, including glucose and heavy metals, potentially broadening their use in diagnostics.
View Article and Find Full Text PDF

Fully integrated and slidable paper-embedded plastic microdevice for point-of-care testing of multiple foodborne pathogens.

Biosens Bioelectron

June 2019

Department of BioNano Technology, Gachon University 1342 Seongnam-daero, Sujeong-gu, Seongnam-si, Gyeonggi-do, 13120, South Korea. Electronic address:

This study presents a slidable paper-embedded plastic microdevice fully integrated with DNA extraction, loop-mediated isothermal amplification (LAMP), and colorimetric detection functionalities. The developed microdevice consists of three layers that allow a sliding movement to mix the sample and reagents for DNA purification, amplification, and detection in a sequential manner. An FTA card was employed in the main chamber for DNA extraction and purification from intact bacterial cells.

View Article and Find Full Text PDF

A fully integrated slidable and valveless microsystem, which performs solid phase DNA extraction (SPE), micro-polymerase chain reaction (μPCR) and micro-capillary electrophoresis (μCE) coupled with a portable genetic analyser, has been developed for forensic genotyping. The use of a slidable chip, in which a 1 μL-volume of the PCR chamber was patterned at the center, does not necessitate any microvalves and tubing systems for fluidic control. The functional micro-units of SPE, μPCR, and μCE were fabricated on a single glass wafer by conventional photolithography, and the integrated microdevice consists of three layers: from top to bottom, a slidable chip, a channel wafer in which a SPE chamber, a mixing microchannel, and a CE microchannel were fabricated, and a Ti/Pt resistance temperature detector (RTD) wafer.

View Article and Find Full Text PDF

A total integrated genetic analysis microsystem was developed, which consisted of solid phase extraction (SPE), polymerase chain reaction (PCR), and immunochromatographic strip (ICS) parts for multiplex colorimetric detection of pathogenic Staphylococcus aureus (S. aureus) and Escherichia coli O157:H7 (E. coli O157:H7) on a portable genetic analyzer.

View Article and Find Full Text PDF