Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background: Cardiac resynchronization therapy (CRT) with multipoint left ventricular (LV) pacing (MultiPoint™ Pacing [MPP], St. Jude Medical) improves acute LV function and LV reverse remodeling at 3 months.

Objective: The purpose of this study was to test the hypothesis that MPP can also improve LV function at 12 months.

Methods: Consecutive patients receiving a CRT implant (Unify Quadra MP™ or Quadra Assura MP™ CRT-D and Quartet™ LV lead, St. Jude Medical) were randomized to receive pressure-volume (PV) loop optimized biventricular pacing with either conventional cardiac resynchronization therapy (CONV) or MPP. CRT response was defined by a reduction in end-systolic volume (ESV) ≥15% relative to BASELINE as determined by a blinded observer and alive status.

Results: Forty-four patients (New York Heart Association class III, ejection fraction [EF] 29% ± 6%, QRS 152 ± 17 ms) were enrolled and randomized to either CONV (N = 22) or MPP (N = 22). During the observation period, 2 patients died of noncardiac causes and 2 patients were lost to follow-up. After 12 months, 12 of 21 patients (57%) in the CONV group and 16 of 21 patients (76%) in the MPP group were classified as CRT responders (P = .33). ESV reduction and EF increase relative to BASELINE were significantly greater with MPP than with CONV (ESV: median -25%, interquartile range [IQR] [-39% to -20%] vs median -18%, IQR [-25% to -2%], P = .03; EF: median +15%, IQR [8% to 20%] vs median +5%, IQR [-1% to 8%], P <.001).

Conclusion: Sustaining the trend observed 3 months postimplant, PV loop-guided multipoint LV pacing resulted in greater LV reverse remodeling and increased LV function at 12 months compared to PV loop-guided conventional CRT.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.hrthm.2015.02.008DOI Listing

Publication Analysis

Top Keywords

cardiac resynchronization
12
resynchronization therapy
12
multipoint left
8
left ventricular
8
ventricular pacing
8
jude medical
8
conv mpp
8
relative baseline
8
patients
6
mpp
5

Similar Publications

Nearly 30% of patients with cardiac resynchronization therapy (CRT) are non- responders. Speckle tracking echocardiography (STE) parameters are able to evaluate electromechanical dyssynchrony that could improve outcomes. We aim to examine the association between various STE parameters with CRT response and clinical outcomes in heart failure patients.

View Article and Find Full Text PDF

Cardiac resynchronization therapy (CRT) improves outcomes in heart failure, but prior interventions like percutaneous mitral annuloplasty may hinder lead placement. We present a 70-year-old male with ischemic cardiomyopathy and severe functional mitral regurgitation who previously received a Carillon device. Due to coronary sinus inaccessibility, left bundle branch area pacing optimized cardiac resynchronization therapy (LOT-CRT) was performed.

View Article and Find Full Text PDF

Background: Genetic aetiologies of early-onset arrhythmias and cardiomyopathy (CM) are common, but timely diagnosis requires a high index of suspicion.

Case Summary: An asymptomatic 47-year-old man presented to cardiology clinic for smartwatch low-rate alarms. His brother had exertional syncope and died in his 20s from heart failure.

View Article and Find Full Text PDF

Background: The prevalence of cardiac implantable electronic devices (CIEDs) in Tunisia is rising because of increased life expectancy and broader indications. This has led to a higher incidence of complications related to vascular access, device pockets, leads, and patient characteristics.

Objective: We aimed to evaluate the prevalence, types, and predictors of complications occurring within the first year after CIED implantation and to profile the demographic and epidemiologic characteristics of CIED recipients in Tunisia.

View Article and Find Full Text PDF