An image-guidance system for dynamic dose calculation in prostate brachytherapy using ultrasound and fluoroscopy.

Med Phys

Department of Electrical and Computer Engineering, Johns Hopkins University, Baltimore, Maryland 21218 and Department of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University, Baltimore, Maryland 21231.

Published: September 2014


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Purpose: Brachytherapy is a standard option of care for prostate cancer patients but may be improved by dynamic dose calculation based on localized seed positions. The American Brachytherapy Society states that the major current limitation of intraoperative treatment planning is the inability to localize the seeds in relation to the prostate. An image-guidance system was therefore developed to localize seeds for dynamic dose calculation.

Methods: The proposed system is based on transrectal ultrasound (TRUS) and mobile C-arm fluoroscopy, while using a simple fiducial with seed-like markers to compute pose from the nonencoded C-arm. Three or more fluoroscopic images and an ultrasound volume are acquired and processed by a pipeline of algorithms: (1) seed segmentation, (2) fiducial detection with pose estimation, (3) seed matching with reconstruction, and (4) fluoroscopy-to-TRUS registration.

Results: The system was evaluated on ten phantom cases, resulting in an overall mean error of 1.3 mm. The system was also tested on 37 patients and each algorithm was evaluated. Seed segmentation resulted in a 1% false negative rate and 2% false positive rate. Fiducial detection with pose estimation resulted in a 98% detection rate. Seed matching with reconstruction had a mean error of 0.4 mm. Fluoroscopy-to-TRUS registration had a mean error of 1.3 mm. Moreover, a comparison of dose calculations between the authors' intraoperative method and an independent postoperative method shows a small difference of 7% and 2% forD90 and V100, respectively. Finally, the system demonstrated the ability to detect cold spots and required a total processing time of approximately 1 min.

Conclusions: The proposed image-guidance system is the first practical approach to dynamic dose calculation, outperforming earlier solutions in terms of robustness, ease of use, and functional completeness.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4149691PMC
http://dx.doi.org/10.1118/1.4893761DOI Listing

Publication Analysis

Top Keywords

dynamic dose
16
image-guidance system
12
dose calculation
12
localize seeds
8
seed segmentation
8
fiducial detection
8
detection pose
8
pose estimation
8
seed matching
8
matching reconstruction
8

Similar Publications

We focused on a paper titled "Radiation with immunotherapy may be a double-edged sword-how can we learn from recent negative clinical trials?", which was published in recently. Herein, we initially provided three complementary viewpoints from biological perspectives involved in the dynamic alterations of the tumor microenvironment, which may contribute to a more comprehensive understanding of the superiority of stereotactic body radiation therapy (SBRT).

View Article and Find Full Text PDF

NSG-SGM3 humanized mouse models are well-suited for studying human immune physiology but are technically challenging and expensive. We previously characterized a simplified NSG-SGM3 mouse, engrafted with human donor CD34 hematopoietic stem cells without receiving prior bone marrow ablation or human secondary lymphoid tissue implantation, that still retains human mast cell- and basophil-dependent passive anaphylaxis responses. Its capacities for human antibody production and human B cell maturation, however, remain unknown.

View Article and Find Full Text PDF

Background: Lung cancer (LC) is the leading cause of cancer death in Ireland, yet no national screening programme exists. While low-dose computed tomography (LDCT) screening reduces lung cancer mortality by approximately 20% in high-risk populations, its cost-effectiveness in Ireland remains uncertain. Evidence on the economic burden of lung cancer care and the feasibility of screening is needed to support policy decisions.

View Article and Find Full Text PDF

Background: Radiotherapy workflows conventionally deliver one treatment plan multiple times throughout the treatment course. Non-coplanar techniques with beam angle optimization or dosimetrically optimized pathfinding (DOP) exploit additional degrees of freedom to improve spatial conformality of the dose distribution compared to widely used techniques like volumetric-modulated arc therapy (VMAT). The temporal dimension of dose delivery can be exploited using multiple plans (sub-plans) within one treatment course.

View Article and Find Full Text PDF

Background: Understanding respiratory motions of liver and its surrogate organs is crucial for precise dose delivery in liver cancer radiotherapy. Although these motions have been studied for respiratory motion management in the supine posture, few studies have quantified them and evaluated their correlations in the upright posture.

Purpose: This study quantified the respiratory motions of liver and surrogate organs and evaluated the correlations between the liver motions and surrogate signals for respiratory motion monitoring in both the supine and upright postures.

View Article and Find Full Text PDF