Article Abstract

Objectives: To compare in vitro biomechanical properties of different designs of epoxy-pin external skeletal fixator (ESF) constructs.

Study Design: Mechanical testing study.

Sample Population: Four epoxy-pin ESF design constructs (uniplanar [EU], multiplanar-I [EM-I], multiplanar-II [EM-II], and circular [EC]) were mechanically tested in compression, bending, and torsion.

Methods: Four different designs of free-form epoxy-pin external fixator constructs were developed using 1.5 mm K-wires and epoxy resin mounted in an ultra-high density polyethylene rod (20 mm diameter). Three-point fixation was done in each fragment, and the distance between fixation wires, and between the rod and the side bars was kept constant in all the designs. A 5 mm gap was maintained at the center of the fixation rod to simulate an unstable fracture condition. The fixator constructs (n = 12 of each design) were subjected to mechanical testing in axial compression, bending, or torsion. Load-deformation curves were generated and mechanical properties were compared between construct types.

Results: EU was the weakest design. Under compression, constructs EM-I, EM-II, and EC were similar. Under bending, EM-I and EM-II had similar strength, whereas EC was strongest. Under torsion, EC was strongest, followed by EM-II, EM-I, and EU; EM-II provided double the rotational stability of EM-I.

Conclusions: Overall, EC followed by EM-II epoxy-pin fixator designs had better mechanical strength.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1532-950X.2014.12128.xDOI Listing

Publication Analysis

Top Keywords

epoxy-pin external
12
em-i em-ii
12
mechanical properties
8
properties designs
8
designs epoxy-pin
8
external skeletal
8
mechanical testing
8
compression bending
8
fixator constructs
8
mechanical
5

Similar Publications

Objective:  The aim of this study was to evaluate epoxy-pin external skeletal fixation technique for the treatment of open fractures in calves and foals.

Study Design:  Twenty-eight calves and four foals (weighing 45-105 kg) with fractures distal to the stifle or elbow made the subject for the retrospective study. The pins (2.

View Article and Find Full Text PDF

Objectives: To compare in vitro biomechanical properties of different designs of epoxy-pin external skeletal fixator (ESF) constructs.

Study Design: Mechanical testing study.

Sample Population: Four epoxy-pin ESF design constructs (uniplanar [EU], multiplanar-I [EM-I], multiplanar-II [EM-II], and circular [EC]) were mechanically tested in compression, bending, and torsion.

View Article and Find Full Text PDF