Assessment of the performance of automated focimeters in the measurement of single vision spectacle lenses.

Clin Exp Optom

Optics & Radiometry Laboratory, School of Optometry and Vision Science, University of New South Wales, Kensington, New South Wales, Australia.

Published: July 2014


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background: The accuracy and precision of any instrument should not be taken for granted. While there is an international standard for checking focimeters, there is no report of any study on their performance.

Method: A sample set of 51 focimeters (11 brands), were used to measure the spherical power of a set of lenses and the prismatic power of two lenses complying with ISO 9342-1:2005 and other calibrated prismatic lenses and the spherical power of some grey filters.

Results: The mean measured spherical power corresponded very closely with the calibrated values; however, the spread of results was substantial and 10 focimeters did not comply with ISO 8598:1996. The measurement of prism was much more accurate and precise and all the focimeters complied easily. With the grey filters, about one-third of the focimeters either showed erratic reading or an error with the equivalent of category 4 sunglasses. On the other hand, nine focimeters had stable and accurate reading on a filter with a luminous transmittance of 0.5 per cent.

Conclusions: These results confirm that, in common with all other measurement instruments, there is a need to ensure that a focimeter is reading accurately and precisely over the range of refractive powers and luminous transmittances. The accurate and precise performance of an automated focimeter over its working life cannot be assumed. Checking before purchase with a set of calibrated lenses and some dark sunglass tints will indicate the suitability of a focimeter. Routine checking with the calibrated lenses will inform the users if a focimeter continues to indicate accurately.

Download full-text PDF

Source
http://dx.doi.org/10.1111/cxo.12132DOI Listing

Publication Analysis

Top Keywords

spherical power
12
performance automated
8
accurate precise
8
calibrated lenses
8
focimeters
7
lenses
6
assessment performance
4
automated focimeters
4
focimeters measurement
4
measurement single
4

Similar Publications

Topic: To compare the outcomes of surgical approaches to correct ametropia following cataract and lens surgery.

Clinical Relevance: Despite advancements in the field of biometry and intraocular lens (IOL) power calculation formulas, complete elimination of refractive surprises following cataract and lens surgery is impossible. Preferred Practice Patterns acknowledges the possibility of refractive surprise following cataract surgery; however, no recommendations regarding the preferred treatment have been given.

View Article and Find Full Text PDF

Plasmonic nanoparticles boost low-current perovskite LEDs governed by photon recycling effects.

RSC Adv

September 2025

Instituto de Ciencia de Materiales de Madrid, ICMM-CSIC C/Sor Juana Inés de la Cruz, 3 Madrid 28049 Spain

Perovskite light-emitting diodes (PeLEDs) have emerged as a promising technology for next-generation display and lighting applications, thanks to their remarkable colour purity, tunability, and ease of fabrication. In this work, we explore the incorporation of plasmonic spherical nanoparticles (NPs) directly embedded into the green-emitting CsPbBr perovskite layer in a PeLED as a strategy to enhance both its optical and electrical properties. We find that plasmonic effects directly boost spontaneous emission while also influencing charge carrier recombination dynamics.

View Article and Find Full Text PDF

Background: Multi-ion radiotherapy using carbon, oxygen, and neon ions aims to improve local control by increasing dose-averaged linear energy transfer (LET) in the target. However, there has been limited understanding of how to utilize variables for multi-ion treatment planning such as the selection and arrangement of ion species.

Purpose: An in silico study was conducted to explore the feasibility of increasing a minimum LET, and the optimal selection and arrangement of ion species in multi-ion therapy for increasing LET in tumors of varying sizes mimicking bone and soft tissue sarcomas (BSTS).

View Article and Find Full Text PDF

Sugarcane () was employed as a sustainable carbon source to synthesize three-dimensional (3D) spherical manganese carbonate (MnCO) microspheres, offering a green route to advanced electrode material for high-energy-density symmetric supercapacitors. Although numerous synthesis strategies and material modifications have been explored, a detailed evaluation of environmentally friendly synthesis pathways remains essential. In this study, MnCO microspheres were successfully synthesized via a sugar-derived green synthesis followed by hydrothermal treatment.

View Article and Find Full Text PDF

Purpose: To evaluate intraocular lens (IOL) power calculation of a non-diffractive extended depth of focus (EDOF) IOL after myopic laser in situ keratomileusis (LASIK) without historical data.

Methods: In this consecutive case series, patients who had undergone lens surgery with implantation of a non-diffractive EDOF IOL after myopic laser in situ keratomileusis (LASIK) at the Department of Ophthalmology, University Hospital Frankfurt, Frankfurt, Germany, were included. Preoperative assessments included biometry and tomography using Scheimpflug technology (Pentacam; Oculus Optikgeräte GmbH).

View Article and Find Full Text PDF