Fixation and mounting of porcine aortic valves for use in mock circuits.

Int J Artif Organs

Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna - Austria and Department of Cardiac Surgery, Medical University of Vienna, Vienna - Austria and Ludwig Boltzmann Cluster for Cardiovascular Research, Vienna - Austria.

Published: October 2013


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Purpose: Investigations of the circulatory system in vitro use mock circuits that require valves to mimic the cardiac situation. Whereas mechanical valves increase water hammer effects due to inherent stiffness and do not allow the use of pressure lines or catheters, bioprosthetic valves are expensive and of limited durability in test fluids. Therefore, we developed a cheap, fast, alternative method to mount valves obtained from the slaughterhouse in mock circuits.

Methods: Porcine aortic roots were obtained from the abattoir and used either in native condition or after fixation. Fixation was performed at a constant retrograde pressure to ensure closed valve position. Fixation time was 4 h in a 0.5%-glutaraldehyde phosphate buffer. The fixed valves were molded into a modular mock circulation connector using a fast curing silicone. Valve functionality was evaluated in a pulsatile setting (cardiac output = 4.7 l/min, heart rate = 80 beats/min) and compared before and after fixation. Leaflet motion was recorded with a high-speed camera and valve insufficiency was quantified by leakage flow under steady pressure application (80 mmHg).

Results: Under physiological conditions the aortic valves showed almost equal leaflet motion in native and fixed conditions. However, the leaflets of the native valves showed lower stiffness and more fluttering during systole than the fixed specimens. Under retrograde pressure, fresh and fixed valves showed small leakage flows of <30 ml/min.

Conclusions: The new mounting and fixation procedure is a fast method to fabricate low cost biologic valves for the use in mock circuits.

Download full-text PDF

Source
http://dx.doi.org/10.5301/ijao.5000230DOI Listing

Publication Analysis

Top Keywords

valves
9
porcine aortic
8
aortic valves
8
mock circuits
8
retrograde pressure
8
fixed valves
8
leaflet motion
8
fixation
5
fixation mounting
4
mounting porcine
4

Similar Publications

ObjectiveTo compare between L pharyngeal flap alone and combined L pharyngeal flap with phenytoin for repair of palatal fistula and velopharyngeal insufficiency (VPI).MethodsTwenty patients with soft palate fistula and persistent VPI who were divided into two groups. In the first group, superiorly based L pharyngeal flap was harvested from the oropharynx and inserted into the soft palate to close the fistula after fistula trimming.

View Article and Find Full Text PDF

Background: Loeys-Dietz syndrome (LDS) is a rare connective tissue disorder (CTD) with musculoskeletal, craniofacial, and cardiovascular features with a prevalence of approximately 1:50,000. Morbidity and mortality often occur earlier in patients with LDS compared to patients with other CTDs.

Case Summary: We present a teenager with subacute heart failure, 4/6 holosystolic murmur with diastolic rumble, facial differences, and arachnodactyly.

View Article and Find Full Text PDF

Background: Transcatheter pulmonary valve implantation (TPVI) has emerged as a viable alternative to surgical pulmonary valve replacement for patients with congenital heart disease and right ventricular outflow tract dysfunction. However, its adoption in low-resource settings has been limited.

Case Summary: We report the first successful TPVI procedures in Tanzania.

View Article and Find Full Text PDF

Comparison of short- and long-term outcomes between transcatheter and surgical aortic valve replacement for bicuspid aortic valve stenosis: a systematic review and meta-analysis.

Int J Surg

September 2025

Department of Cardiovascular Surgery, The Affiliated Hospital, Southwest Medical University, Metabolic Vascular Diseases Key Laboratory of Sichuan Province, Key Laboratory of Cardiovascular Remodeling and Dysfunction, Luzhou, Sichuan, PR China.

Objective: This meta-analysis aimed to compare the perioperative safety and efficacy of transcatheter aortic valve replacement (TAVR) versus surgical aortic valve replacement (SAVR) in bicuspid aortic valve (BAV) stenosis.

Methods: We systematically analyzed studies from PubMed, Cochrane Library, Embase, and CNKI comparing TAVR and SAVR in BAV stenosis. Outcomes included postoperative mortality, complications, all-cause survival, and freedom from stroke.

View Article and Find Full Text PDF

Epidemiology, pathophysiology, diagnosis and management of atrial functional mitral regurgitation: An expert opinion paper.

ESC Heart Fail

September 2025

Institute of Cardiology, ASST Spedali Civili, Department of Medical and Surgical Specialties, Radiological Sciences, and Public Health, University of Brescia, Brescia, Italy.

Atrial functional mitral regurgitation (AFMR) is an increasingly recognized subtype of mitral regurgitation, characterized by left atrial remodelling and mitral annular dilation in the absence of primary mitral valve disease or left ventricular dysfunction. Closely linked to chronic atrial fibrillation and heart failure with preserved ejection fraction, AFMR is associated with poor clinical outcomes and represents a growing therapeutic challenge. This expert opinion paper summarizes current evidence on the epidemiology, pathophysiology, diagnosis and management strategies, including medical therapy and emerging data supporting surgical and transcatheter interventions in selected patients.

View Article and Find Full Text PDF