Article Abstract

Background And Purpose: The robotic monopolar scissors tip cover accessory (TCA) is an insulation device that prevents current leak from surfaces on the instrument other than the scissors tip. Reports of insulation failure and patient injury have been made but not systematically studied. We investigate the incidence of TCA failure and potential mechanisms that lead to malfunction.

Materials And Methods: Eighty TCAs (40 first and 40 second generation) were collected after a single urologic or gynecologic surgery at a single academic institution. Each TCA was inspected for damage under a microscope. Current leak was measured in an electrolysis solution, and electrical arcing was tested in a porcine kidney model. Log-rank and Friedman two-way analysis of variance by ranks compared failure at different angulations and power settings. Chi-square analysis compared failure between first and second generation TCAs (f-TCAs and s-TCAs).

Results: Visible insulation defects were detected in 39% of f-TCAs (size range <0.5-2.75 mm). Electrical arcing was observed in 33% of f-TCAs. Arcing increased with greater wrist angulation (P=0.014) and higher power settings (P=0.048). Minor damage was observed in 25% of s-TCAs. No electrical arcing was observed in any of the s-TCAs despite angulation and power strain. s-TCAs demonstrated significantly less failure than f-TCAs (P<0.001).

Conclusion: In this study, 33% of f-TCAs demonstrated insulation failure after a single surgical use. f-TCA failure increased with greater wrist angulation and electrical power. The improved design of s-TCAs demonstrated no current leak or arcing on ex-vivo testing and appears to have significantly decreased failure potential. Centers still using the f-TCA should transition to the s-TCA and be vigilant in reporting device failures to prevent unnecessary patient morbidity.

Download full-text PDF

Source
http://dx.doi.org/10.1089/end.2013.0052DOI Listing

Publication Analysis

Top Keywords

cover accessory
8
current leak
8
second generation
8
compared failure
8
failure
5
prospective analysis
4
analysis robotic
4
robotic cover
4
accessory failure
4
failure background
4

Similar Publications

Bacteriophages are the most abundant entities on earth and exhibit vast genetic and phenotypic diversity. Exploitation of this largely unexplored molecular space requires identification and functional characterization of genes that act at the phage-host interface. So far, this has been restricted to few model phage-host systems that are amenable to genetic manipulation.

View Article and Find Full Text PDF

Background: Menthol cigarette alternative products (e.g., menthol accessories, non-menthol cooling cigarettes) have entered the market and may bypass menthol cigarette restrictions.

View Article and Find Full Text PDF

Focal Cortical Dysplasia (FCD) is a common cause of drug-resistant epilepsy. These abnormalities arise during embryonic development and are challenging to classify due to their complex nature. The most recent classification update of FCD incorporates genetic and epigenetic results with other clinical data for the management of epilepsy associated with these lesions.

View Article and Find Full Text PDF

Preemptive aneurysm sac embolization for infrarenal endovascular aneurysm repair.

J Vasc Surg

August 2025

Division of Cardiac, Thoracic, and Vascular Surgery, New York-Presbyterian, Columbia University Irving Medical Center, Columbia University Vagelos College of Physicians and Surgeons, New York, NY. Electronic address:

Objective: Aneurysm sac behavior has been associated with long-term survival and reinterventions, spurring an interest in more active management of the aneurysm sac during endovascular aortic repair (EVAR). We therefore investigated the utility of sac embolization with shape memory polymer (SMP) plugs (Shape Memory Medical), a novel, biodegradable, nonartifact-producing implant inserted into the aneurysm sac between the graft and vessel wall to promote sac thrombosis and regression.

Methods: We retrospectively studied all patients undergoing EVAR of infrarenal aneurysms at two centers from February 2022, to January 2024, where SMP plugs were used.

View Article and Find Full Text PDF

A framework for personal protective equipment use in laboratories: regulatory compliance and employee protection.

Front Public Health

August 2025

Office of Occupational Safety and Health, U.S. Food and Drug Administration, U.S. Department of Health and Human Services, Washington, DC, United States.

The Occupational Safety and Health Administration (OSHA) Personal Protective Equipment (PPE) Standard, outlined in 29 CFR 1910 Subpart I states that Protective gear - including personal protective clothing, respiratory equipment, and barriers or shields- must be supplied, properly used, and kept clean and functional whenever there are hazards present. These hazards may arise from processes, environmental conditions, chemicals, radiation, or mechanical sources that could potentially cause injury or harm bodily functions through skin contact, inhalation, or absorption. Employers must supply PPE to their employees as determined by hazard assessments of the workplace and duties performed by employees.

View Article and Find Full Text PDF