Objectives: The Vicorder is a new brachial cuff-based device that estimates central blood pressure (cBP) using a brachial-to-aortic transfer function. The aim of this study was to evaluate cBP estimated by the Vicorder.

Methods: During cardiac angiography, cBP estimated by the Vicorder and the SphygmoCor was evaluated against simultaneous invasive cBP in 50 patients. The two devices were also compared noninvasively in a separate group of 90 healthy individuals.

Results: Central SBP (cSBP) obtained with each device satisfied the American Association for the Advancement of Medical Instrumentation accuracy criteria when peripheral waveforms were calibrated to invasive mean arterial pressure (MAP)/DBP: estimated - invasive cSBP difference, -4.0 ± 7.4 mmHg, Vicorder, P < 0.001; -1.4 ± 7.9 mmHg, SphygmoCor, P = 0.21. When oscillometric brachial SBP/DBP was used for peripheral waveform calibration, cSBP was underestimated by Vicorder (Δ -6.4 ± 7.4 mmHg, P < 0.001 versus invasive) and more so by SphygmoCor (Δ -11.9 ± 7.2 mmHg, P < 0.001 versus invasive). Conversely, cSBP was more closely estimated by SphygmoCor when waveforms were calibrated to brachial MAP/DBP (Δ -2.8 ± 9.4 mmHg, P = 0.04 versus invasive). In the noninvasive study, Vicorder cSBP correlated well with SphygmoCor cSBP when SphygmoCor waveforms were calibrated to brachial MAP/DBP (121 ± 16 versus 121 ± 17 mmHg, P = 0.2) but not when brachial SBP/DBP was used for calibration (115 ± 19 mmHg, P < 0.001).

Conclusion: The Vicorder and SphygmoCor devices provide reliable estimates of cSBP when calibrated to invasive pressure. When calibrated to brachial BP, both devices underestimated cSBP, although this was attenuated when SphygmoCor was calibrated to brachial MAP/DBP. Vicorder may be a simple alternative to tonometry-based methods for noninvasive assessment of cBP.

Download full-text PDF

Source
http://dx.doi.org/10.1097/HJH.0b013e32835a8eb1DOI Listing

Publication Analysis

Top Keywords

cuff-based device
8
central blood
8
blood pressure
8
cbp estimated
8
evaluation vicorder
4
vicorder novel
4
novel cuff-based
4
device noninvasive
4
noninvasive estimation
4
estimation central
4

Similar Publications

Accuracy of a wearable watch-type oscillometric blood pressure monitor in rest and ambulatory blood pressure measurements.

Hypertens Res

August 2025

Department of Cardiovascular Medicine, State Key Laboratory of Medical Genomics, Shanghai Key Laboratory of Hypertension, The Shanghai Institute of Hypertension, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China.

The aim of this study was to evaluate the accuracy of a watch-type oscillometric BP monitor HUAWEI WATCH D2 for rest and ambulatory BP measurements and compare with a conventional ambulatory BP monitor (A&D TM-2430, Tokyo, Japan). Subjects were recruited to fulfill the age, gender, BP and cuff distributions of the AAMI/ESH/ISO Universal Standard in the general population. Two cuffs of the test device were used for wrist circumference 13-16 (medium) and 16-21 cm (large), respectively.

View Article and Find Full Text PDF

Underestimation of systolic pressure in cuff-based blood pressure measurement.

PNAS Nexus

August 2025

Department of Engineering, University of Cambridge, Trumpington Street, Cambridge CB2 1PZ, United Kingdom.

High blood pressure (hypertension) is the number one risk factor for premature death. Hypertension is asymptomatic, so blood pressure must be regularly monitored to diagnose it. In auscultatory blood pressure measurement, a patient's systolic (maximum) and diastolic (minimum) blood pressure are inferred from the pressure in an inflatable cuff wrapped around the arm.

View Article and Find Full Text PDF

Introduction: Hypertension is the leading risk factor for cardiovascular disorders. Early detection and initiation of treatment have been identified as the most effective ways to reduce the burden of hypertension. The most common method for detecting hypertension is blood pressure measurement, typically performed with cuff-based devices, where systolic pressure (SBP) and diastolic pressure (DBP) are measured through Korotkoff sounds.

View Article and Find Full Text PDF

Aims: Cardiovascular diseases are a global health crisis, with hypertension as a significant risk factor. Traditional cuff-based blood pressure measurements have various limitations, prompting the exploration of photoplethysmography as an alternative for continuous monitoring. This study aimed to assess a cuff-calibrated wrist-worn photoplethysmography-based blood pressure device against European Society of Hypertension recommendations.

View Article and Find Full Text PDF

Aims: Hypertension is a leading cause of death worldwide, yet many hypertensive cases remain undiagnosed. Wearable, cuffless blood pressure (BP) monitors could be deployed at scale, but their accuracy remains undetermined.

Methods And Results: This study validated a popular consumer-grade wearable BP monitor (W-BPM, Aktiia), using a medical-grade ambulatory device (A-BPM, Mobil-O-Graph), for reference.

View Article and Find Full Text PDF