Background: In cardiac SPECT perfusion imaging, motion correction of the data is critical to the minimization of motion introduced artifacts in the reconstructed images. Software-based (data-driven) motion correction techniques are the most convenient and economical approaches to fulfill this purpose. However, the accuracy is significantly affected by how the data complexities, such as activity overlap, non-uniform tissue attenuation, and noise are handled.

Methods: We developed STASYS, a new, fully automatic technique, for motion detection and correction in cardiac SPECT. We evaluated the performance of STASYS by comparing its effectiveness of motion correcting patient studies with the current industry standard software (Cedars-Sinai MoCo) through blind readings by two readers independently.

Results: For 204 patient studies from multiple clinical sites, the first reader identified (1) 69 studies with medium to large axial motion, of which STASYS perfectly or significantly corrected 86.9% and MoCo 72.5%; and (2) 20 studies with medium to large lateral motion, of which STASYS perfectly or significantly corrected 80.0% and MoCo 60.0%. The second reader identified (1) 84 studies with medium to large axial motion, of which STASYS perfectly or significantly corrected 82.2% and MoCo 76.2%; and (2) 34 studies with medium to large lateral motion, of which STASYS perfectly or significantly corrected 58.9% and MoCo 50.0%.

Conclusions: We developed a fully automatic software-based motion correction technique, STASYS, for cardiac SPECT. Clinical studies showed that STASYS was effective and corrected a larger percent of cardiac SPECT studies than the current industrial standard software.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2708325PMC
http://dx.doi.org/10.1007/s12350-009-9096-7DOI Listing

Publication Analysis

Top Keywords

cardiac spect
20
studies medium
16
medium large
16
motion stasys
16
stasys perfectly
16
perfectly corrected
16
fully automatic
12
motion correction
12
motion
11
motion detection
8

Similar Publications

The development of pulmonary hypertension (PH) after splenectomy is one of the recently controversial issues. This study aims to investigate whether splenectomy itself is an independent risk factor for the development of PH or if the primary contributor to PH development is the underlying condition that necessitated splenectomy. This study was conducted prospectively.

View Article and Find Full Text PDF

Purpose: Cardiac noradrenergic denervation visualized by meta-[I]iodobenzylguanidine ([I]MIBG) imaging supports the diagnosis of Parkinson's disease (PD). Recently, meta-[F] fluorobenzylguanidine ([F]MFBG) PET demonstrated favorable imaging characteristics compared with [I]MIBG scintigraphy for neuroendocrine tumors. We assessed [F]MFBG dosimetry and myocardial pharmacokinetics in healthy controls and PD patients.

View Article and Find Full Text PDF

Introduction: Targeted infection imaging is crucial for accurate diagnosis in postpartum women. This project uses 99mTc-labeled cefixime to develop a radiopharmaceutical for detecting, distinguishing, and treating infections and abscesses in women.

Method: Technetium (TcO4-) chelated with cefixime, reduced by stannous chloride, confirmed via thin-layer chromatography.

View Article and Find Full Text PDF

Background: Cardiac positron emission tomography (PET) and single-photon emission computed tomography (SPECT) are widely used for the assessment of coronary artery disease. While SPECT remains more available, workforce shortages and training demands contribute to geographic disparities in PET availability, impacting patient access to advanced imaging. Therefore, we assessed trends in the U.

View Article and Find Full Text PDF

Systemic amyloidosis is a complex disorder, making early and accurate diagnosis challenging. The most common types are associated with misfolded transthyretin or immunoglobulin light chains, where cardiac and renal amyloidosis portend the worst prognosis. Peptide p5+14 can bind all types of amyloid via multivalent electrostatic interactions.

View Article and Find Full Text PDF