98%
921
2 minutes
20
Background: Ensuring reference material (RM) commutability is crucial for evaluating measurement traceability in order to standardize laboratory tests. However, commutability assessment is not routinely performed. We assessed whether RMs prepared following CLSI C37-A guidelines could be used without assessing commutability by evaluating their commutability for four lipid measurements using the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) and CLSI EP14 protocols.
Methods: We analyzed total cholesterol (TC), triglycerides (TGs), high-density lipoprotein cholesterol (HDL-C), and low-density lipoprotein cholesterol (LDL-C) in frozen sera from 20 individuals and 11 RMs, prepared by the Korea Disease Control and Prevention Agency-Laboratory Standardization Project (per CLSI C37-A), using six routine measurement procedures (MPs). Regression equations and 95% prediction intervals derived from single-donor sera were analyzed following CLSI EP14. The IFCC protocol was used to assess differences in inter-MP biases between RM and clinical samples. The effect of the TG concentration on commutability was evaluated by analyzing biases between MP results and reference procedure-assigned values.
Results: RMs were commutable for most MP pairs for TC and TG. Commutability for HDL-C and LDL-C varied across RMs, with RM10 and RM11 showing higher TG levels (2.38 and 2.95 mmol/L, respectively) and lower commutability. Increased bias percentages from assigned values were observed for RMs with higher TG levels.
Conclusions: RMs prepared per CLSI C37-A were commutable with most MP pairs for TC and TG. Elevated TG levels affected HDL-C and LDL-C commutability, highlighting the need to consider TG concentrations during RM preparation and assess commutability to standardize laboratory tests.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.3343/alm.2025.0010 | DOI Listing |
Ann Lab Med
September 2025
Department of Laboratory Medicine, Konkuk University Medical Center, Konkuk University School of Medicine, Seoul, Korea.
Background: Ensuring reference material (RM) commutability is crucial for evaluating measurement traceability in order to standardize laboratory tests. However, commutability assessment is not routinely performed. We assessed whether RMs prepared following CLSI C37-A guidelines could be used without assessing commutability by evaluating their commutability for four lipid measurements using the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) and CLSI EP14 protocols.
View Article and Find Full Text PDFClin Chem Lab Med
September 2023
Division of Laboratory Sciences, National Center for Environmental Health, Centers for Disease Control and Prevention, Atlanta, GA, USA.
Objectives: The clinical use of soluble transferrin receptor (sTfR) as an iron status indicator is hindered by a lack of assay standardization and common reference ranges and decision thresholds. In 2009, the WHO and National Institute for Biological Standards and Controls (NIBSC) released a sTfR reference material (RM), 07/202, for assay standardization; however, a comprehensive, formal commutability study was not conducted.
Methods: This study evaluated the commutability of WHO 07/202 sTfR RM and human serum pools and the impacts of their use as common calibrators.
Clin Chem Lab Med
February 2020
Department of Pathology, Virginia Commonwealth University, Richmond, VA, USA.
Manufacturers of in vitro diagnostic medical devices, clinical laboratories, research laboratories and calibration laboratories require commutable reference materials that can be used in the calibration hierarchies of medical laboratory measurement procedures used for human specimens to establish metrological traceability to higher order reference systems. Commutable materials are also useful in external quality assessment surveys. In order to achieve these goals, matrix-based reference materials with long-term stability, appropriate measurand concentrations and commutability with individual human specimens are required.
View Article and Find Full Text PDF