Severity: Warning
Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 197
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1075
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3195
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 317
Function: require_once
98%
921
2 minutes
20
Background: Urinary catecholamine metabolites are well-established biomarkers for neuroblastoma (NB). Homovanillic acid (HVA) and vanillylmandelic acid (VMA) are the most frequently measured metabolites within SIOPEN - Catecholamine Working Group laboratories. Here, we evaluated the performance of a new LC-MS/MS in vitro diagnostic (IVD) kit for HVA and VMA to facilitate inter-laboratory harmonization.
Methods: HVA and VMA and their deuterated internal standards were analyzed with a commercial method, on a ThermoFisher Quantiva LC-MS/MS. Validation was performed first using internal quality control and external quality assessment (IQC and EQA) samples. Next by clinical validation on 120 samples, previously tested by HPLC-ECD. Finally, 36 samples were exchanged between SIOPEN reference laboratories and analyzed by three methods.
Results: Using QCs and EQA the method was validated in a wide calibration range (4.61-830 µmol/L for HVA and 4.44-800 µmol/L for VMA). Intra-day CVs ( = 5) were 7 and 8 % for HVA and 5 and 6 % for VMA for QC low and QC high, respectively; Inter-day CV% were 7 and 3 % for HVA and 2 and 7 % for VMA at QC low and QC high, respectively. Its application to 120 clinical samples confirmed a high diagnostic accuracy. The inter-laboratory quality control assessment showed interchangeable results ( 0,73 and 0.15 for HVA and VMA, respectively).
Conclusion: The LC-MS/MS IVD method could be considered a useful tool for clinical laboratories involved in the measurement of catecholamines, contributing to harmonization efforts.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC12105103 | PMC |
http://dx.doi.org/10.1016/j.jmsacl.2025.04.007 | DOI Listing |