Article Abstract

Objectives: This study compared the clinical accuracy of two different stationary face scanners, employing progressive capture and multi-view simultaneous capture scanning technologies.

Methods: Forty dentate volunteers participated in the study. Soft tissue landmarks were marked with a pen on the participants' faces to measure the distances between them. Clinical measurements were manually obtained using a digital vernier caliper by two independent examiners. The participants were then scanned using one of two stationary face scanners: Obiscanner (Fifthingenium), which employs progressive capture technology requiring the subject's head to rotate during image acquisition, or RAYFace (RAY), which utilizes multiple cameras to simultaneously capture a complete 3D image. The scans were imported into mesh-processing software, and digital measurements were taken by the same examiners. Data analysis included pairwise comparison tests and the calculation of the intra-class correlation coefficient (ICC; α = 0.05).

Results: Digital measurements were significantly longer than clinical measurements across all measured distances (p < 0.001). Comparisons between the scanners revealed that vertical measurements using RAYface exhibited greater percentage differences compared to those using Obiscanner (p < 0.05), while horizontal measurements were more variable with Obiscanner than those obtained using RAYface (p < 0.05). Intra-examiner differences were significant for both methods (p < 0.001), although inter-examiner differences were only significant for clinical measurements (p < 0.001), not for digital measurements (p > 0.05). Inter-examiner reliability for digital measurements was high (ICC≥0.99).

Conclusions: Significant differences were observed in the accuracy of the two stationary face scanners using progressive capture and multi-view simultaneous capture scanning technologies, with each device demonstrating specific strengths and limitations.

Clinical Significance: Although face scanners offer relatively high accuracy and consistency, particularly across different acquisition technologies, careful consideration of their performance characteristics is essential for optimizing accuracy in facial measurements.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jdent.2024.105553DOI Listing

Publication Analysis

Top Keywords

face scanners
20
digital measurements
16
stationary face
12
progressive capture
12
clinical measurements
12
measurements
10
scanning technologies
8
accuracy stationary
8
capture multi-view
8
multi-view simultaneous
8

Similar Publications

This article introduces a digital technique for virtual articulator mounting by employing the scan of a facebow worn by the patient as a virtual reference. The digital technique enables the transfer of the maxillary arch orientation relative to the cranial base into a CAD-CAM environment (Ceramill Mind; AmannGirrbach), without the need for ionizing radiation or identification of facial landmarks. By digitally aligning the intraoral scans of the dental arches (Trios 4; 3Shape) with a 3D facial scan and the scanned facebow in position (Artex; AmannGirrbach), clinicians can reproduce the cranium-to-maxilla spatial relationship accurately and intuitively.

View Article and Find Full Text PDF

Smartphone-based 3D facial scanning applications in orthodontic diagnosis: A cross-sectional study.

J Dent Res Dent Clin Dent Prospects

June 2025

Department of Orthodontics, Saveetha Dental College and Hospitals, Saveetha Institute of Medical and Technical Sciences, Saveetha University, Chennai, India.

Background: Innovations in smartphone technology have transformed diagnostics in healthcare. In orthodontics, these applications can serve as potential diagnostic tools. The present study analyzed the accuracy of two smartphone-based face scanners compared to a standard CBCT-based facial scanner and evaluated scanning duration, user experience, and preferences for orthodontic diagnostics.

View Article and Find Full Text PDF

Objective: This in vitro study evaluated mesh density and 3D accuracy of three facial scanning systems: a stationary scanner, a handheld scanner, and a mobile scanning application.

Methods: A mannequin head with 14 fixed landmarks across three facial regions served as the master model. An industrial scanner (ATOS Q, Zeiss Co.

View Article and Find Full Text PDF

Statement Of Problem: Various digital workflows are available to capture the maxillomandibular relationship at an increased vertical dimension, including the use of intraoral scanners (IOSs) and optical jaw tracking systems. However, limited evidence exists for the virtual determination of an increased vertical dimension using jaw-tracking devices.

Purpose: The purpose of this clinical study was to compare the acquisition of the maxillomandibular relationship at an increased vertical dimension with the virtual determination through a jaw-tracking system and by using an IOS.

View Article and Find Full Text PDF

Flexible ultrathin metasurface with open-ring spiral resonators for SNR enhancement in low-field 0.2 T MRI.

Magn Reson Imaging

August 2025

Institute of Electrical Engineering Chinese Academy of Sciences, Beijing, China, ; University of Chinese Academy of Sciences, Beijing, China. Electronic address:

Low-field magnetic resonance imaging (MRI) systems (<0.5 T) have emerged as increasingly valuable tools for emergency and bedside imaging due to their cost-effectiveness and portability, but their adoption is limited by low signal-to-noise ratio (SNR). Conventional optimization methods offer limited improvements, while current metasurface technologies face challenges related to clinical integration, such as excessive dimensionality, and inadequate SNR enhancement at low magnetic field strengths.

View Article and Find Full Text PDF