Article Abstract

Background: Chronic venous insufficiency (CVI) is a common disease with a high prevalence. Incompetent venous valves are considered as one of the main causes. Besides compression therapy, various surgical therapies are practiced, whereby the reconstruction of valves is of central importance. There is an unmet clinical need, no valve prosthesis is commercially available to date. This work introduces two versions of a patented prosthetic bicuspid valve design made of electrospun thermoplastic silicone polycarbonate polyurethane (TSPCU) nanofiber leaflets attached in a nitinol stent, and their performance in static and pulsatile operation.

Results: The valves mainly fulfill the requirements widely accepted in literature. Valves of both versions were functional in the physiological pressure range up to 50 mmHg with design specific differences.

Conclusions: The here introduced design versions act as a platform technology and can be tailored for an intended implantation site. Evaluation of the original and modified valve concept demonstrated efficacy, with limitations at higher loads for original design. At the current state, the modification is preferable for fabrication, as one processing step is eliminated. Moreover, specific design recommendations could be drawn for valves of similar basic structure. Future work will focus on long-term performance and biocompatibility prior to the initiation of preclinical in vivo studies.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11606079PMC
http://dx.doi.org/10.1186/s12938-024-01316-xDOI Listing

Publication Analysis

Top Keywords

valves
5
design
5
transcatheter bicuspid
4
bicuspid venous
4
valve
4
venous valve
4
valve prostheses
4
prostheses fluid
4
fluid mechanical
4
mechanical performance
4

Similar Publications

[Diagnosis of non-autoimmune hemolysis in the adult].

Rev Med Interne

September 2025

Service d'hématologie biologique, CHU d'Amiens-Picardie, Amiens, France; HEMATIM UR4666, université Picardie Jules-Verne, Amiens, France.

The diagnosis of hemolysis is still based on straightforward biochemical parameters: haptoglobin (the most sensitive), lactate dehydrogenase (LDH), and unconjugated bilirubin. Anemia is not always present. Reticulocyte counts typically exceed 120×10/L, except in cases of associated vitamin deficiency or during the very early phase of acute hemolysis.

View Article and Find Full Text PDF

A left ventricular sub-mitral thrombus without an aneurysm is a previously unreported rare occurrence. We aim to bring attention to this finding in a case of colorectal adenocarcinoma.An early 60s-year-old female presented with bleeding per rectum, weight loss and fatigue and was found to have colorectal carcinoma with metastasis based on examination, imaging and biopsy findings.

View Article and Find Full Text PDF

Moderate Secondary Mitral Regurgitation: Evolving Evidence and Management Strategies.

J Am Coll Cardiol

September 2025

Institute of Cardiology, ASST Spedali Civili, Department of Medical and Surgical Specialties, Radiologic Sciences and Public Health, University of Brescia, Brescia, Italy. Electronic address:

Secondary mitral regurgitation (SMR) is common in patients with heart failure (HF). Although randomized clinical trials have been focused on the treatment of severe SMR, the prognostic role and potential for treatment of moderate SMR cannot be overlooked. The randomized RESHAPE-HF2 trial included patients with moderate and severe SMR with consistent findings in both groups, raising the hypothesis that transcatheter correction of moderate SMR could have beneficial effects, although this needs further investigation.

View Article and Find Full Text PDF