Article Abstract

(MC), a member of the complex, can cause infections in patients after open-heart surgery due to contaminated heater-cooler units (HCUs). The transmission route of HCU-related MC infection is non-inhalational, and infection can occur in patients without previously known immune deficiency. Patients may develop endocarditis of the prosthetic valve, infection of the vascular graft, and/or manifestations of disseminated mycobacterial infection (splenomegaly, arthritis, hepatitis, nephritis, myocarditis, etc.). MC infections have serious outcomes (30-50% recurrence rate, 20-67% mortality rate). In 2015, an international outbreak of infections among patients undergoing cardiothoracic surgeries was associated with exposure to contaminated LivaNova 3T HCUs (formerly Stöckert 3T heater-cooler system, London, United Kingdom). In response to the global outbreak, many international agencies have issued directives and recommendations in order to reduce the risk of MC infection in cardiac surgery. Whole-genome sequencing (WGS) technology can be used to describe the global spread and dynamics of MC infections, to characterize local outbreaks, and also to identify sources of infection in hospital settings. In order to minimize the risk of contamination of HCUs and reduce the risk of patient infection, it is imperative that healthcare facilities establish a program of regular cleaning and disinfection maintenance procedures as well as monitoring of the water used and the air in the operating room, in accordance with the manufacturer's procedure.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11434768PMC
http://dx.doi.org/10.3390/pathogens13090781DOI Listing

Publication Analysis

Top Keywords

heater-cooler units
8
infection
8
infections patients
8
reduce risk
8
spread heater-cooler
4
units infection
4
risk
4
infection risk
4
risk heart
4
heart surgery
4

Similar Publications

This article presents an image dataset of non-tuberculous mycobacteria (NTM) isolated from heater-cooler units (HCU) using the recently developed NTM Elite agar (bioMérieux, Marcy-l'Étoile, France). NTM (e.g.

View Article and Find Full Text PDF

Objective: To enumerate and describe the effect of whole genome sequencing (WGS) on epidemiological investigations of healthcare-associated transmission of nontuberculous mycobacteria (NTM).

Design: Systematic review.

Methods: We performed a literature search using targeted search terms to identify articles meeting inclusion criteria.

View Article and Find Full Text PDF

Mycobacterium chimaera, belonging to the Mycobacterium avium complex, is an opportunistic environmental mycobacterium which has been isolated from medical device water samples such as Heater Cooler Units (HCU). Laboratories currently use culture-based diagnostic methods to detect M. chimaera, but these take a long time to obtain results.

View Article and Find Full Text PDF
Article Synopsis
  • The study assesses the effectiveness of the BD BACTEC™ MGIT™ system for detecting non-tuberculous mycobacteria (NTM) in sanitary water, focusing on samples from heater-cooler units and haemodialysis instruments over a five-year period.
  • A total of 1,369 water samples from HCUs and 747 from dialysis machines were tested, with 261 and 20 cultures yielding NTM, respectively, verified through advanced techniques.
  • Results show that the MGIT system is not only reliable and reproducible for clinical applications but also effective for environmental testing of NTM, detecting concentrations as low as 4 CFU/mL.
View Article and Find Full Text PDF
Article Synopsis
  • The study monitored 29 Maquet heater-cooler units (HCUs) over a 63-month period to track the presence of non-tuberculous mycobacteria (NTM) in response to the replacement of LivaNova devices due to contamination concerns.
  • Results showed a significant increase in NTM positivity, particularly after three years, with 100% of HCU40 and 62% of HU35 devices colonized, primarily by Mycobacterium gordonae.
  • The findings indicate that current disinfection methods using chloramine-T are insufficient, highlighting the need for improved device safety designs and new disinfection protocols.
View Article and Find Full Text PDF