Article Abstract

Introduction: New dialysis membranes with new properties are being developed to improve efficacy and tolerance. The hemocompatibility of a polymeric biomaterial is influenced by the layer of water at the blood membrane interface. The new dialyzer TORAY NV-U has a membrane Hydrolink™, designed to suppress platelet adhesion and to improve the hemocompatibility. Until now, there is no experience in online hemodiafiltration (OL-HDF).The objective of the present study is to evaluate the efficacy of this new membrane in OL-HDF therapy compared to another membrane commonly used. Other objectives are to evaluate the inflammatory response, hemodynamic tolerance, and the anticoagulation regimes.

Methods: This is a prospective pilot study performed in five anuric patients receiving OL-HDF. For 1 month patients were kept with their usual dialyzer FX1000 (FMC). Subsequently, the dialyzer was changed to TORAY NV-U (Hydrolink) for 1 month. In the last dialysis session of each dialyzer, blood tests were performed to evaluate inflammation and depurative capacity.

Results: We did not find differences in medium size removal molecules and convective volume: FX1000: 31 ± 9 l per session and Hydrolink™ 30 ± 8 l;  = 0.7); β2microglobulin reduction ratio (RR) FX1000 FMC 83 ± 3%; Hydrolink™ 79 ± 4;  = 0.14; Myoglobin RR FX1000 FMC 72 ± 7%; Hydrolink™ 76 ± 4;  = 0.28. We did not find differences in inflammation parameters: serum IL6 with FX1000 6.0 ± 4.2 pg/mL; Hydrolink™ 7.6 ± 5.0 pg/mL;  = 0.3.During all sessions with the two dialyzers there was adequate plasmatic filling, reaching 85 % filling. All patients had "good" dialyzer status in all dialysis sessions with TORAY NV-U, while the dialyzer status with FX1000 was "good" in 20% of the sessions, "medium" in 30%, and "dirty" in the remaining 50% dialysis sessions.

Conclusions: The new dialyzer Hydrolink™, TORAY NV-U is not inferior to perform OL-HDF compared to dialyzers usually used for this therapy, and could allow decrease heparin doses. Further studies with a bigger sample size and longer follow-up will answer if Hydrolink improves inflammation and assess a better hemodynamic tolerance.

Download full-text PDF

Source
http://dx.doi.org/10.1177/03913988241274735DOI Listing

Publication Analysis

Top Keywords

toray nv-u
20
fx1000 fmc
12
dialyzer
8
dialyzer toray
8
online hemodiafiltration
8
hemodynamic tolerance
8
find differences
8
dialyzer status
8
hydrolink™
7
fx1000
6

Similar Publications

Introduction: New dialysis membranes with new properties are being developed to improve efficacy and tolerance. The hemocompatibility of a polymeric biomaterial is influenced by the layer of water at the blood membrane interface. The new dialyzer TORAY NV-U has a membrane Hydrolink™, designed to suppress platelet adhesion and to improve the hemocompatibility.

View Article and Find Full Text PDF

Purpose:: Microaggregates have often been observed during hemodialysis and are clearly associated with complications of hemodialysis therapy. In this study, we aimed to clarify the effects of two polysulfone membranes, with different abilities to activate blood cells, on the formation of these microaggregates; we also investigated their molecular mechanisms.

Methods:: Human whole blood was circulated through a mini-module dialyzer using the membranes in vitro; platelet-neutrophil complexes in blood were determined by flow cytometry.

View Article and Find Full Text PDF

Activation of blood cells during hemodialysis is considered to be a significant determinant of biocompatibility of the hemodialysis membrane because it may affect patient health adversely through microvascular inflammation and oxidative stress. This study found very different cell activation among various polysulfone (PSf) hemodialysis membranes. For example, CX-U, a conventional PSf membrane, induced marked adhesion of platelets to its surface and increased surface expression of activated CD11b and production of reactive oxygen species (ROS) by neutrophils; while NV-U, a hydrophilic polymer-immobilized PSf membrane, caused little platelet adhesion and slight CD11b expression and ROS production by neutrophils.

View Article and Find Full Text PDF