Article Abstract

Introduction: This study was designed to examine the capacity of intraoperative optical coherence tomography (OCT) to predict the postimplant position of the glaucoma drainage device PreserfloTM.

Methods: 13 eyes (mean age 65.42 (14.89) years) underwent PreserfloTM (Santen, Osaka, Japan) placement. Before surgery, participants were subjected to a comprehensive ophthalmic examination (intraocular pressure (IOP), cup to disk ratio (C/D), visual field, OCT, endothelial cell count). Anterior segment OCT scans were obtained intraoperatively using a Rescan 700 OCT system (Carl Zeiss Meditec, Inc., Oberkochen, Germany). One day postsurgery, anterior segment OCT using the Spectralis OCT (Heidelberg Engineering GmbH) was performed in a sitting position to capture the same chamber cross-section as before. The main outcome variables were tube-endothelium distance (T-E) and tube length (TL) in the anterior chamber measured using both OCT systems. Correlation between intraoperative and office measurements was examined through Pearson correlation (r) and intraclass correlation coefficients (ICC).

Results: Mean intraoperative and in-office T-E were 625.26 (SD 366.60) versus 561.16 (SD 364.62) µm respectively ( = 0.540). Intraoperative and in-office anterior chamber TL were 1386 (SD 701.82) and 1433.91 (SD 713.55) µm, respectively ( = 0.029). Excellent correlation was observed between both sets of T-E ( = 0.992;  = 0.008) and TL ( = 0.984;  = 0.016) values. Both OCT systems showed good agreement yielding ICCs of 0.992 ( < 0.001) for T-E and 0.995 ( = 0.001) for TL.

Discussion: Excellent correlation was observed between our intraoperative and postoperative OCT measurements. These results support the usefulness of intraoperative OCT to confirm the correct position of an implanted PreserfloTM microshunt.

Download full-text PDF

Source
http://dx.doi.org/10.1177/11206721241272169DOI Listing

Publication Analysis

Top Keywords

postimplant position
8
oct
8
anterior segment
8
segment oct
8
anterior chamber
8
oct systems
8
intraoperative in-office
8
intraoperative
5
intraoperative oct
4
oct check
4

Similar Publications

Background: Cardiac resynchronization therapy (CRT) has been established as a key component in the management of patients with heart failure (HF) with reduced ejection fraction in addition to pharmacologic therapy. Several automatic algorithms have been developed to optimize the timing cycle settings in CRT, especially AV delay which was associated with improvement of the response to CRT. The present study aims to investigate whether the novel device-based SyncAV algorithm could elicit better synchrony and acute hemodynamic response.

View Article and Find Full Text PDF

: Obstructive sleep apnea (OSA) is a prevalent disorder characterized by airway collapse during sleep. Continuous positive airway pressure (CPAP) is first-line treatment but adherence can decay over time due to intolerance. Hypoglossal nerve stimulation (HNS) has emerged as an alternative, especially for CPAP-intolerant patients.

View Article and Find Full Text PDF

The transcatheter EVOQUE tricuspid valve replacement system is an approved device for the treatment of patients with symptomatic severe tricuspid regurgitation despite medical therapy both in the United States and in Europe. Specific imaging requirements are needed in the screening and in the intraprocedural phase as well as to assess this valve after implantation, in order to assess its positioning and performance. In this review, focus will be on the postoperative echocardiographic assessment of device positioning and stability; evaluation of residual central tricuspid regurgitation or paravalvular leak; valve thrombosis/degeneration; interaction of the device frame with the right ventricle and adjacent structures, including the aortic valve and interventricular septum; and placement of pacemaker leads.

View Article and Find Full Text PDF

Accurate localization of the epileptogenic zone (EZ) is essential for surgical success in temporal lobe epilepsy. While stereoelectroencephalography (SEEG) and structural magnetic resonance imaging (MRI) provide complementary insights, existing unimodal methods fail to fully capture epileptogenic brain activity, and multimodal fusion remains challenging due to data complexity and surgeon-dependent interpretations. To address these issues, we proposed a novel multimodal framework to improve EZ localization with SEEG-drived electrophysiology with structural connectivity in temporal lobe epilepsy.

View Article and Find Full Text PDF

Background: Extra-vascular implantable cardioverter defibrillator (EV-ICD) is an emerging technique for ICD implantation with substernal lead placement. The system avoids the vasculature and can deliver anti-tachycardia and pause prevention pacing, with a generator size smaller than the existing subcutaneous device. While this device is now widely available, little is known about potential complications that may follow its implantation in real life context.

View Article and Find Full Text PDF