Severity: Warning
Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 197
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3165
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 317
Function: require_once
98%
921
2 minutes
20
Purpose: A daily quality assurance (QA) check in proton therapy is ensuring that the range of each proton beam energy in water is accurate to 1 mm. This is important for ensuring that the tumor is adequately irradiated while minimizing damage to surrounding healthy tissue. It is also important to verify the total charge collected against the beam model. This work proposes a time-efficient method for verifying the range and total charge of proton beams at different energies using a multilayer Faraday collector (MLFC).
Methods: We used an MLFC-128-250 MeV comprising 128 layers of thin copper foils separated by thin insulating Kapton layers. Protons passing through the collector induce a charge on the metallic foils, which is integrated and measured by a multichannel electrometer. The charge deposition on the foils provides information about the beam range.
Results: Our results show that the proton beam range obtained using MLFC correlates closely with the range obtained from commissioning water tank measurements for all proton energies. Upon applying a range calibration factor, the maximum deviation is 0.4 g/cm. The MLFC range showed no dependence on the number of monitor units and the source-to-surface distance. Range measurements collected over multiple weeks exhibited stability. The total charge collected agrees closely with the theoretical charge from the treatment planning system beam model for low- and mid-range energies.
Conclusions: We have calibrated and commissioned the use of the MLFC to easily verify range and total charge of proton beams. This tool will improve the workflow efficiency of the proton QA.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11320672 | PMC |
http://dx.doi.org/10.1177/15330338241262610 | DOI Listing |