Purpose: To compare the spectral performance of two different DSCT (DSCT-Pulse and DSCT-Force) on virtual monoenergetic images (VMIs) at low energy levels.

Methods: An image quality phantom was scanned on the two DSCTs at three dose levels: 11/6/1.8 mGy. Level 3 of an advanced modeled iterative reconstruction algorithm was used. Noise power spectrum and task-based transfer function were computed on VMIs from 40 to 70 keV to assess noise magnitude and noise texture (f) and spatial resolution (f). A detectability index (d') was computed to assess the detection of one contrast-enhanced abdominal lesion as a function of the keV level used.

Results: For all dose levels and all energy levels, noise magnitude was significantly higher (p < 0.05) with DSCT-Pulse than with DSCT-Force (12.6 ± 2.7 % at 1.8 mGy, 9.1 ± 2.9 % at 6 mGy and 4.0 ± 2.7 % at 11 mGy). For all energy levels, f values were significantly higher (p < 0.05) with DSCT-Pulse than with DSCT-Force at 1.8 mGy (4.8 ± 3.9 %) and at 6 mGy (5.5 ± 2.5 %) but similar at 11 mGy (0.2 ± 3.6 %; p = 0.518). For all energy levels, f values were significantly higher with DSCT-Pulse than with DSCT-Force (12.7 ± 5.6 % at 1.8 mGy, 17.9 ± 4.5 % at 6 mGy and 13.1 ± 2.6 % at 11 mGy). For all keV, similar d' values were found with both DSCT-Force and DSCT-Pulse at 11 mGy (-1.0 ± 3.1 %; p = 0.084). For other dose levels, d' values were significantly lower with DSCT-Pulse than with DSCT-Force (9.1 ± 3.2 % at 1.8 mGy and -6.3 ± 3.9 % at 6 mGy).

Conclusion: Compared with the DSCT-Force, the DSCT-Pulse improved noise texture and spatial resolution, but noise magnitude was slightly higher and detectability slightly lower, particularly when the dose level was reduced.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejmp.2024.103429DOI Listing

Publication Analysis

Top Keywords

spectral performance
8
virtual monoenergetic
8
monoenergetic images
8
dose levels
8
noise magnitude
8
comparison spectral
4
performance dual-source
4
dual-source systems
4
systems low-energy
4
low-energy virtual
4

Similar Publications

Accurate differentiation between persistent vegetative state (PVS) and minimally conscious state and estimation of recovery likelihood in patients in PVS are crucial. This study analyzed electroencephalography (EEG) metrics to investigate their relationship with consciousness improvements in patients in PVS and developed a machine learning prediction model. We retrospectively evaluated 19 patients in PVS, categorizing them into two groups: those with improved consciousness ( = 7) and those without improvement ( = 12).

View Article and Find Full Text PDF

Development and characterization of a prototype selenium-75 high dose rate brachytherapy source.

Med Phys

September 2025

Medical Physics Unit, Department of Oncology, Faculty of Medicine, McGill University, Montréal, Québec, Canada.

Background: Se ( 120 days, 215 keV) offers advantages over Ir ( 74 days, 360 keV) as a high dose rate brachytherapy source due to its lower gamma energy and longer half-life. Despite its widespread use in industrial gamma radiography, a Se brachytherapy source has yet to be manufactured.

Purpose: A novel Se-based source design with a vanadium diselenide core, titled the SeCure source, was proposed.

View Article and Find Full Text PDF

Machine Learning-Aided Screening and Design Rule Discovery for LWIR-Transparent Optical Materials.

J Chem Inf Model

September 2025

Department of Chemistry and Biochemistry, University of Arizona, Tucson, Arizona 85721-0041, United States.

The development of low-cost, high-performance materials with enhanced transparency in the long-wavelength infrared (LWIR) region (800-1250 cm/8-12.5 μm) is essential for advancing thermal imaging and sensing technologies. Traditional LWIR optics rely on costly inorganic materials, limiting their broader deployment.

View Article and Find Full Text PDF

An ongoing goal of top-down mass spectrometry is to increase the performance for larger proteins. Using higher energy activation methods, like 193 nm ultraviolet photodissociation (UVPD), offers the potential to cause more extensive fragmentation of large proteins and thereby yield greater sequence coverage. Obtaining high sequence coverage requires confident identification and assignment of fragment ions, and this process is hampered by spectral congestion and low signal-to-noise ratio (S/N) of the fragment ions.

View Article and Find Full Text PDF

Congenital hearing loss is a significant health problem, with a worldwide incidence of >6 per 1000 live births. Late diagnosis will delay appropriate treatment, leading to potential neurodevelopment problems. Early diagnosis requires neonatal hearing screening, where one of the most used techniques is automated Auditory Brainstem Responses (aABR).

View Article and Find Full Text PDF