Severity: Warning
Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 197
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3165
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 317
Function: require_once
98%
921
2 minutes
20
Background: The purpose of this study was to perform a biomechanical analysis to compare different medial column fixation patterns for valgus pilon fractures in a case-based model.
Methods: Based on the fracture mapping, 48 valgus pilon fracture models were produced and assigned into four groups with different medial column fixation patterns: no fixation (NF), K-wires (KW), intramedullary screws (IS), and locking compression plate (LCP). Each group contained wedge-in and wedge-out subgroups. After fixing each specimen on the machine, gradually increased axial compressive loads were applied with a load speed of one millimeter per minute. The maximum peak force was set at 1500 N. Load-displacement curves were generated and the axial stiffness was calculated. Five different loads of 200 N, 400 N, 600 N, 800 N, 1000 N were selected for analysis. The specimen failure was defined as resultant loading displacement over 3 mm.
Results: For the wedge-out models, Group-IS showed less displacement (p < 0.001), higher axial stiffness (p < 0.01), and higher load to failure (p < 0.001) than Group-NF. Group-KW showed comparable displacement under loads of 200 N, 400 N and 600 N with both Group-IS and Group-LCP. For the wedge-in models, no statistical differences in displacement, axial stiffness, or load to failure were observed among the four groups. Overall, wedge-out models exhibited less axial stiffness than wedge-in models (all p < 0.01).
Conclusions: Functional reduction with stable fixation of the medial column is essential for the biomechanical stability of valgus pilon fractures and medial column fixation provides the enough biomechanical stability for this kind of fracture in the combination of anterolateral fixation. In detail, the K-wires can provide a provisional stability at an early stage. Intramedullary screws are strong enough to provide the medial column stability as a definitive fixation. In future, this technique can be recommended for medial column fixation as a complement for holistic stability in high-energy valgus pilon fractures.