Article Abstract

Purpose: Quantitative MRI techniques such as T2 mapping are useful in comprehensive evaluation of various pathologies of the knee joint yet require separate scans to conventional morphological measurements and long acquisition times. The recently introduced 3D MIXTURE (Multi-Interleaved X-prepared Turbo-Spin Echo with Intuitive Relaxometry) technique can obtain simultaneous morphologic and quantitative information of the knee joint. To compare MIXTURE with conventional methods and to identify differences in morphological and quantitative information.

Methods: Phantom studies were conducted, and in vivo human scans were performed (20 patients) presented with knee arthralgia. MIXTURE is based on 3D TSE without and with T2 preparation modules in an interleaved manner for both morphology with PDW and fat suppressed T2W imaging as well as quantitative T2 mapping within one single scan. Image quality and lesion depiction were visually assessed and compared between MIXTURE and conventional 2D TSE by two experienced radiologists. Contrast-to-noise ratio was used to assess the adjacent tissue contrast in a quantitative way for both obtained PDW and fat suppressed T2W images. Quantitative T2 values were measured in phantom and from in vivo knee cartilage.

Results: The overall diagnostic confidence and contrast-to-noise ratio were deemed comparable between MIXTURE and 2D TSE. While the chosen T2 preparation modules for MIXTURE rendered consistent T2 values comparing to the current standard, measured cartilage T2 values ranged from 26.1 to 50.7 ms, with significant difference between the lesion and normal areas (p < 0.05).

Conclusions: MIXTURE can help to provide high-resolution information for both anatomical and pathological assessment.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejrad.2024.111579DOI Listing

Publication Analysis

Top Keywords

multi-interleaved x-prepared
8
intuitive relaxometry
8
knee joint
8
mixture conventional
8
preparation modules
8
pdw fat
8
fat suppressed
8
suppressed t2w
8
contrast-to-noise ratio
8
mixture
7

Similar Publications

Purpose: Quantitative MRI techniques such as T2 mapping are useful in comprehensive evaluation of various pathologies of the knee joint yet require separate scans to conventional morphological measurements and long acquisition times. The recently introduced 3D MIXTURE (Multi-Interleaved X-prepared Turbo-Spin Echo with Intuitive Relaxometry) technique can obtain simultaneous morphologic and quantitative information of the knee joint. To compare MIXTURE with conventional methods and to identify differences in morphological and quantitative information.

View Article and Find Full Text PDF
Article Synopsis
  • The study evaluated a new imaging technique called MIXTURE, used for assessing cartilage and meniscus in knee specimens, compared to traditional turbo spin-echo (TSE) sequences.
  • Standardized cartilage defects were created in human cadaver knees, and both MIXTURE and TSE sequences were analyzed for their effectiveness in showing defects and measuring relaxation times.
  • The results showed that while both imaging methods had comparable visual results, MIXTURE provided a time-efficient way to obtain useful quantitative data on cartilage conditions.
View Article and Find Full Text PDF

Quantitative MRI techniques such as T2 and T1ρ mapping are beneficial in evaluating knee joint pathologies; however, long acquisition times limit their clinical adoption. MIXTURE (Multi-Interleaved X-prepared Turbo Spin-Echo with IntUitive RElaxometry) provides a versatile turbo spin-echo (TSE) platform for simultaneous morphologic and quantitative joint imaging. Two MIXTURE sequences were designed along clinical requirements: "MIX1", combining proton density (PD)-weighted fat-saturated (FS) images and T2 mapping (acquisition time: 4:59 min), and "MIX2", combining T1-weighted images and T1ρ mapping (6:38 min).

View Article and Find Full Text PDF
Article Synopsis
  • The study evaluates a new imaging technique called MIXTURE, developed by Philips, which produces high-resolution T2 maps for diagnosing cervical radiculopathy.
  • It compares T2 relaxation times and their variability using MIXTURE with another method, SHINKEI-Quant (S-Q), in both healthy participants and a patient with cervical disc herniation.
  • Results show that MIXTURE provides significantly shorter T2 relaxation times and lower variability in measurements, indicating its potential for more accurate preoperative functional diagnosis of cervical nerve issues.
View Article and Find Full Text PDF