A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 197

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3165
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 317
Function: require_once

How to verify the analytical and clinical performance of ELISA immunoanalysis in the real laboratory practice. PCSK9 as an example. | LitMetric

How to verify the analytical and clinical performance of ELISA immunoanalysis in the real laboratory practice. PCSK9 as an example.

J Immunol Methods

Institute for Clinical and Experimental Medicine, Department of Laboratory Methods, Vídeňská 1958/9, 140 21 Praha 4, Czech Republic; Third Faculty of Medicine, Charles University, Ruská 87, 100 00 Praha 10, Czech Republic.

Published: July 2024


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background: Manufacturers and diagnostic companies often recommend on-site verification of analytical performance in the clinical laboratory. The validation process used by manufacturers is rarely described in detail, and certain information on analytical performance is missing from the product sheet, especially for immunoanalytical methods. We describe an approach to the detailed validation of an ELISA method for the measurement of proprotein convertase subtilisin/kexin type 9 (PCSK9) plasma concentrations. We compared manufacturers' claims of analytical performance with data obtained in the field laboratory using several approaches.

Methods: We used the Human Proprotein Convertase 9/PCSK9 Quantikine ELISA diagnostic kit (R&D systems, Bio-Techne Ltd., Abingdon Science Park, Abingdon, UK) and three levels of quality control solution Quantikine Immunoassay Control Group 235 (R&D systems, Bio-Techne Ltd., Abingdon Science Park, Abingdon, UK) to verify precision. We measured the concentration of PCSK9 using the DS2 ELISA Reader (Dynex Technologies GmbH, Denkendorf, Germany). We used analysis of variance (ANOVA) and the R statistical package (R core team, version 1.4.5). Statistical analysis and terminology were performed according to protocol CLSI EP15-A3, and the reference interval was checked according to CLSI/IFCC C28-A3c.

Results: We found a significant difference between the manufacturer's claims of analytical performance and real data measured in the routine clinical laboratory. The calculated CV (%) for repeatability (calculated by simple estimation of the mean and SD, as used by the manufacturer) was between 5.5% and 7.4%, but the manufacturer's claim was between 4.1% and 6.5%. Using ANOVA, the true repeatability was between 5.0% and 6.9%. Similarly, ANOVA revealed values of CV (%) for within-laboratory imprecision between 6.5% and 9.1%, while the manufacturer's claims were between 4.1% and 5.9%. The recovery ranged from 105.5% to 121.8%. The manufacturer's recommended reference interval was checked and we didn't find any significant difference between men and women.

Conclusions: We describe a comprehensive approach to verify the analytical performance of an ELISA method using the measurement of PCSK9 plasma concentration as an example. We found differences between the results of this approach based on the CLSI EP15-A3 protocol and data provided by the manufacturer. We recommend the verification of analytical performance by more complex statistical tools in laboratory practice.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jim.2024.113693DOI Listing