A Modular Communicative Leadless Pacing-Defibrillator System.

N Engl J Med

From the Department of Cardiology, Amsterdam University Medical Center, Amsterdam (R.E.K., L.V.A.B.), and the Department of Cardiology, St Antonius Ziekenhuis, Nieuwegein (L.V.A.B.) - both in the Netherlands; Emory University Section of Cardiac Electrophysiology, Atlanta (M.S.L., F.M.M.); University

Published: October 2024


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background: The subcutaneous implantable cardioverter-defibrillator (ICD) is associated with fewer lead-related complications than a transvenous ICD; however, the subcutaneous ICD cannot provide bradycardia and antitachycardia pacing. Whether a modular pacing-defibrillator system comprising a leadless pacemaker in wireless communication with a subcutaneous ICD to provide antitachycardia and bradycardia pacing is safe remains unknown.

Methods: We conducted a multinational, single-group study that enrolled patients at risk for sudden death from ventricular arrhythmias and followed them for 6 months after implantation of a modular pacemaker-defibrillator system. The safety end point was freedom from leadless pacemaker-related major complications, evaluated against a performance goal of 86%. The two primary performance end points were successful communication between the pacemaker and the ICD (performance goal, 88%) and a pacing threshold of up to 2.0 V at a 0.4-msec pulse width (performance goal, 80%).

Results: We enrolled 293 patients, 162 of whom were in the 6-month end-point cohort and 151 of whom completed the 6-month follow-up period. The mean age of the patients was 60 years, 16.7% were women, and the mean (±SD) left ventricular ejection fraction was 33.1±12.6%. The percentage of patients who were free from leadless pacemaker-related major complications was 97.5%, which exceeded the prespecified performance goal. Wireless-device communication was successful in 98.8% of communication tests, which exceeded the prespecified goal. Of 151 patients, 147 (97.4%) had pacing thresholds of 2.0 V or less, which exceeded the prespecified goal. The percentage of episodes of arrhythmia that were successfully terminated by antitachycardia pacing was 61.3%, and there were no episodes for which antitachycardia pacing was not delivered owing to communication failure. Of 162 patients, 8 died (4.9%); none of the deaths were deemed to be related to arrhythmias or the implantation procedure.

Conclusions: The leadless pacemaker in wireless communication with a subcutaneous ICD exceeded performance goals for freedom from major complications related to the leadless pacemaker, for communication between the leadless pacemaker and subcutaneous ICD, and for the percentage of patients with a pacing threshold up to 2.0 V at a 0.4-msec pulse width at 6 months. (Funded by Boston Scientific; MODULAR ATP ClinicalTrials.gov NCT04798768.).

Download full-text PDF

Source
http://dx.doi.org/10.1056/NEJMoa2401807DOI Listing

Publication Analysis

Top Keywords

subcutaneous icd
16
leadless pacemaker
16
performance goal
16
antitachycardia pacing
12
major complications
12
exceeded prespecified
12
pacing-defibrillator system
8
icd provide
8
pacemaker wireless
8
wireless communication
8

Similar Publications

Despite the transformative clinical impact of programmed cell death ligand 1/programmed death 1 (PD-1/PD-L1) blockade in hepatocellular carcinoma (HCC), therapeutic efficacy remains limited by the tumor immunosuppressive microenvironment (TIME), with objective response rates persistently below 20 %. To address this critical clinical challenge, we engineered ultrasound (US)-responsive lipid nanobubbles (NBs) co-encapsulating microRNA (miR)-195-5p and shikonin (SK) (designated miR-195-5p/SK-NBs), a dual-functional platform designed to synergize PD-L1 suppression with immunogenic cell death (ICD). The NBs exhibited tumor-selective accumulation through passive and active targeting mechanisms while maintaining biosafety.

View Article and Find Full Text PDF

Background: Implantable cardioverter-defibrillators (ICDs) prevent sudden cardiac death due to ventricular arrhythmia. A novel extravascular ICD (EV-ICD) system provides improved functionality over previous transvenous (TV-ICD) and subcutaneous (S-ICD) alternatives, particularly in younger patients. This includes limited bradycardia pacing, anti-tachycardia pacing therapy, and lower energy defibrillation, all within a smaller device profile compared to the S-ICD.

View Article and Find Full Text PDF

Introduction: Sense-B-noise is a rare cause of inappropriate shock (IAS) of subcutaneous implantable cardioverter defibrillator (S-ICD), which is an oversensing of the vectors using the Sense B and a factor of noncardiac signal oversensing. High-amplitude signals in Sense-B-noise have not been reported previously.

Methods: We present a case of a 47-year-old man, who experienced IAS of S-ICD due to sense-B-noise followed by intense noise of primary and alternate vectors.

View Article and Find Full Text PDF

Introduction: CIED patients are routinely advised against physical activity with a risk of collision because of expected damage to the implanted device. However, no data support this practice.

Methods: Sixteen CIED systems (6 pacemakers, 6 ICD, 4 CRT-D) from all manufacturers were implanted subcutaneously in a porcine thorax and increasing weights were dropped on the experimental setting.

View Article and Find Full Text PDF

Background: A subcutaneous implantable cardioverter defibrillator (S-ICD) is an alternative to a conventional transvenous implantable cardioverter defibrillator for preventing sudden cardiac death. Although posterior chest S-ICD implantation has been recommended for better defibrillation outcomes, little is known about the optimal S-ICD positioning for R-wave detection. Herein, we report two cases of S-ICD recipients in whom antero-inferior chest positioning improved R-wave detection after posterior chest positioning failed.

View Article and Find Full Text PDF