A missing coronary guidewire mimicking aortic dissection-a case report.

Eur Heart J Case Rep

Cardiology Department, Meir Medical Center, Tchernichovsky St 59, Kfar Saba 4418001, Israel.

Published: April 2024


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background: A retained coronary guidewire following coronary angiography is an extremely rare complication. We present a case of a retained coronary guidewire from a percutaneous coronary intervention done 2 years ago.

Case Summary: An 80-year-old asymptomatic man with a history of ischemic heart disease and moderate aortic stenosis presented to the echocardiography lab for routine follow-up. Transthoracic echocardiography showed Moderate aortic stenosis and a suspected linear echogenic structure in the ascending aorta. trans-esophageal echocardiography was performed to reveal a mobile and linear echogenic structure originating from the sinuses of Valsalva/Sinotubular junction and extending to the ascending aorta. An electrocardiogram gated cardiac computed tomography was performed and showed A linear well-defined structure originating from the ostium of the left main coronary artery and extending to the ascending aorta-a coronary guidewire from an earlier procedure. A second look at the last invasive coronary angiography record demonstrated the same finding. A multidisciplinary heart team discussion was obtained and concluded that the risk of surgical or endovascular intervention outweighed the potential benefit. The patient was discharged home for a close clinical and echocardiographic follow-up.

Discussion: A retained coronary guidewire is a rare complication that operators should be aware of. Management should be case-specific depending on clinical presentation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10977952PMC
http://dx.doi.org/10.1093/ehjcr/ytae100DOI Listing

Publication Analysis

Top Keywords

coronary guidewire
20
retained coronary
12
coronary
8
coronary angiography
8
rare complication
8
moderate aortic
8
aortic stenosis
8
linear echogenic
8
echogenic structure
8
ascending aorta
8

Similar Publications

Background: Umbilical arterial catheterisation is a common intervention performed in the neonatal intensive care unit (NICU) especially in extremely preterm and extremely low birth weight neonates. Rarely catheter fracture or breakage can occur, leaving behind part of the catheter in the aorta. A handful of cases have been reported in the literature, with the majority being managed surgically.

View Article and Find Full Text PDF

Robot-assisted percutaneous coronary intervention: a prospective, multicenter, randomized controlled, non-inferiority clinical trial.

J Geriatr Cardiol

August 2025

Department of Cardiology, Beijing Anzhen Hospital, Capital Medical University, Beijing Institute of Heart Lung and Blood Vessel Disease, Beijing Key Laboratory of Precision Medicine of Coronary Atherosclerotic Disease, Clinical Center for Coronary Heart Disease, Capital Medical University, Beijing,

Objective: To evaluate the safety and effectiveness of robot-assisted percutaneous coronary intervention (R-PCI) compared to traditional manual percutaneous coronary intervention (M-PCI).

Methods: This prospective, multicenter, randomized controlled, non-inferior clinical trial enrolled patients with coronary heart disease who met the inclusion criteria and had indications for elective percutaneous coronary intervention. Participants were randomly assigned to either the R-PCI group or the M-PCI group.

View Article and Find Full Text PDF

One of the relatively common anatomical variants of coronary vessels that is often overlooked in clinical practice is coronary artery tortuosity (CAT). CAT can have a significant impact on coronary blood flow and procedural outcomes during percutaneous coronary intervention (PCI). It is defined by bends, curves, or loops within the coronary vasculature that can lead to increased vascular resistance.

View Article and Find Full Text PDF

Background: Guide extension catheters are specially designed for percutaneous coronary intervention (PCI) to enhance backup support of the guide catheter by providing coaxial alignment, thereby allowing deep intubation of the catheter. We have developed an innovative auxiliary support technique utilizing a dual guide extension catheter system, designed to enhance safety and facilitate deep coronary artery access.

Case Summary: A male in his sixties who presented with chest pain was diagnosed with non-ST elevation myocardial infarction.

View Article and Find Full Text PDF

Background: We report the first in-literature animal experiment to validate the intracoronary ECG signal acquired from a coronary wire compared with the direct signal from an epicardial electrode.

Methods: An animal model study was performed in a 40 kg pig. Acute myocardial ischemia was induced by intracoronary balloon inflation for 60 s.

View Article and Find Full Text PDF