Article Abstract

Background: The incorporation of magnetic resonance (MR) imaging in radiotherapy (RT) workflows improves contouring precision, yet it introduces geometrical uncertainties when registered with computed tomography (CT) scans. Synthetic CT (sCT) images could minimize these uncertainties and streamline the RT workflow. This study aims to compare the contouring capabilities of sCT images with conventional CT-based/MR-assisted RT workflows, with an emphasis on managing artefacts caused by surgical fixation devices (SFDs).

Methods: The study comprised a commissioning cohort of 100 patients with cranial tumors treated using a conventional CT-based/MR-assisted RT workflow and a validation cohort of 30 patients with grade IV glioblastomas treated using an MR-only workflow. A CE-marked artificial-intelligence-based sCT product was utilized. The delineation accuracy comparison was performed using dice similarity coefficient (DSC) and average Hausdorff distance (AHD). Artefacts within the commissioning cohort were visually inspected, classified and an estimation of thickness was derived using Hausdorff distance (HD). For the validation cohort, boolean operators were used to extract artefact volumes adjacent to the target and contrasted to the planning treatment volume.

Results: The combination of high DSC (0.94) and low AHD (0.04 mm) indicates equal target delineation capacity between sCT images and conventional CT scans. However, the results for organs at risk delineation were less consistent, likely because of voxel size differences between sCT images and CT scans and absence of standardized delineation routines. Artefacts observed in sCT images appeared as enhancements of cranial bone. When close to the target, they could affect its definition. Therefore, in the validation cohort the clinical target volume (CTV) was expanded towards the bone by 3.5 mm, as estimated by HD analysis. Subsequent analysis on cone-beam CT scans showed that the CTV adjustment was enough to provide acceptable target coverage.

Conclusion: The tested sCT product performed on par with conventional CT in terms of contouring capability. Additionally, this study provides both the first comprehensive classification of metal artefacts on a sCT product and a novel method to assess the clinical impact of artefacts caused by SFDs on target delineation. This methodology encourages similar analysis for other sCT products.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11320867PMC
http://dx.doi.org/10.1186/s13014-024-02422-9DOI Listing

Publication Analysis

Top Keywords

sct images
20
validation cohort
12
sct product
12
sct
9
images conventional
8
conventional ct-based/mr-assisted
8
artefacts caused
8
commissioning cohort
8
hausdorff distance
8
target delineation
8

Similar Publications

Background: This study evaluates the quality of synthetic computed tomography (sCT) images for MR-only radiotherapy in prostate cancer using gamma analysis. A software tool, MRGamma, was developed to address challenges like the absence of electron density maps and registration uncertainties between magnetic resonance imaging (MRI) and planning CT (pCT).

Materials And Methods: Aplication developed in MATLAB assesses Hounsfield units (HU) discrepancies between CT and sCT images via 2D and 3D gamma analysis (GA).

View Article and Find Full Text PDF

Background: Shield-triggered autoinjectors (AIs) aim to reduce needle phobia and accidental needlestick injuries and improve usability. However, they may cause deeper injections due to tissue compression. This study investigates the mechanical response of AI application into abdominal tissue in-vivo.

View Article and Find Full Text PDF

Clinical Relevance: Understanding early microvascular alterations in the retinal and choroidal structures of patients with diabetes mellitus is essential for the timely identification of high-risk individuals and the prevention of vision-threatening complications.

Background: This study evaluates the impact of diabetes mellitus (DM) on retinal and choroidal vascular structures using optical coherence tomography angiography (OCTA) in pre-retinopathic patients. It also explores the relationship between OCTA parameters and HbA1c levels, a marker of long-term glycaemic control.

View Article and Find Full Text PDF

Background: To evaluate the precision of automated segmentation facilitated by deep learning (DL) and dose calculation in adaptive radiotherapy (ART) for nasopharyngeal cancer (NPC), leveraging synthetic CT (sCT) images derived from cone-beam CT (CBCT) scans on a conventional C-arm linac.

Materials And Methods: Sixteen NPC patients undergoing a two-phase offline ART were analyzed retrospectively. The initial (pCT) and adaptive (pCT) CT scans served as gold standard alongside weekly acquired CBCT scans.

View Article and Find Full Text PDF

Kidney allograft rejection occurs in clinically stable patients, but its long-term significance in children is unknown. Previous studies demonstrated that subclinical inflammation is associated with an increased risk of rejection. However, the prevalence and significance of subclinical antibody-mediated rejection and the impact of subclinical (SC) rejection phenotypes on graft survival remains to be assessed.

View Article and Find Full Text PDF