Article Abstract

Background: The delivery of neuroendovascular devices requires a robust proximal access platform. This demand has previously been met with a 6Fr long sheath (8Fr guide) that is placed in the proximal internal carotid artery (ICA) or vertebral artery segments. We share our experience with the first 0.088 inch 8Fr guide catheter designed for direct intracranial access.

Methods: We retrospectively reviewed a prospectively maintained IRB-approved institutional database of the senior authors to identify all cases where the TracStar Large Distal Platform (LDP) was positioned within the intracranial vasculature, defined as within or distal to the petrous ICA, vertebral artery (V3) segments, or transverse sinus. Technical success was defined as safe placement of the TracStar LDP within or distal to the described distal vessel segments with subsequent complication-free device implantation.

Results: Over the 41-month study period from January 2020 to June 2023, 125 consecutive cases were identified in whom the TracStar LDP was navigated into the intracranial vasculature for triaxial delivery of large devices, 0.027 inch microcatheter and greater, for aneurysm treatment (n=108, 86%), intracranial angioplasty/stenting (n=15, 12%), and venous sinus stenting (n=2, 1.6%). All cases used a direct select catheter technique for initial guide placement (no exchange). Posterior circulation treatments occurred in 14.4% (n=18) of cases. Technical success was achieved in 100% of cases. No vessel dissections occurred in any cases.

Conclusion: The TracStar LDP is an 0.088 inch 8Fr guide catheter that can establish direct intracranial access with an acceptable safety profile. This can be achieved in a wide range of neurointerventional cases with a high rate of technical success.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11671879PMC
http://dx.doi.org/10.1136/jnis-2023-021054DOI Listing

Publication Analysis

Top Keywords

tracstar ldp
16
8fr guide
12
technical success
12
intracranial access
8
delivery large
8
ica vertebral
8
vertebral artery
8
artery segments
8
0088 inch
8
inch 8fr
8

Similar Publications

Background: The delivery of neuroendovascular devices requires a robust proximal access platform. This demand has previously been met with a 6Fr long sheath (8Fr guide) that is placed in the proximal internal carotid artery (ICA) or vertebral artery segments. We share our experience with the first 0.

View Article and Find Full Text PDF

Introduction: Venous sinus stenting is a well established alternative to cerebrospinal fluid diversion for the treatment of idiopathic intracranial hypertension (IIH) with associated venous sinus stenosis. During this procedure, distal guide catheter placement within the venous sinuses may be desirable to facilitate stent delivery. We report our initial experience using the TracStar LDP™ (Imperative Care, Campbell, USA, 0.

View Article and Find Full Text PDF

Background: A number of large bore guide catheters are currently available for use in neuroendovascular surgery. This study represents a multi-institutional retrospective series of patients undergoing mechanical thrombectomy with the use of a TracStar Large Distal Platform (LDP) guide catheter and assessed its performance in vivo in 107 patients.

Objective: To review a multi-institutional initial experience with the TracStar LDP guide catheter during mechanical thrombectomy for emergent large vessel occlusion (ELVO).

View Article and Find Full Text PDF

Introduction: The delivery of flow-diverting stents (FDS) necessitates a degree of catheter support beyond that required for endovascular coiling. The TracStar Large Distal Platform (LDP) is a novel 0.088″ platform intended for navigation into the intracranial internal carotid artery (ICA).

View Article and Find Full Text PDF

Background: Most conventional 0.088 inch guide catheters cannot safely navigate intracranial vasculature. The objective of this study is to evaluate the safety of stroke thrombectomy using a novel 0.

View Article and Find Full Text PDF