A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 197

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3165
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 317
Function: require_once

Comparison of a coupling system and the suture method in end-to-side microvascular anastomosis in head and neck reconstruction. | LitMetric

Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background: Use of coupling devices in microvascular anastomosis continues to increase, but it is not yet actively used in end-to-side (ETS) anastomosis because there is no standard method. Therefore, we propose an easy and time-saving ETS micro-anastomosis method using coupling devices in head and neck reconstruction and compare it with the conventional suture method.

Materials And Methods: We retrospectively reviewed 30 consecutive cases (43 anastomoses) of ETS anastomosis in head and neck reconstruction from 2018 to 2022. Patient characteristics, operative details, and anastomosis time were evaluated. When using the coupling device, a cross incision was created in the recipient vessel to form vascular flaps. By pulling the flaps out of the ring, the intact vessel lining was fixed. Other procedures were the same as for a traditional anastomosis.

Results: The mean anastomosis time was 12.81 min (range, 6.7-24.87) for the suture and 4.96 min (range, 2.02-9.4) for the coupling device, a statistically significant difference (p-value <.00005). There was no venous insufficiency or flap failure with either method.

Conclusions: ETS venous anastomosis using the coupling device is an easy-to-use, safe, and timesaving procedure for head and neck reconstruction.

Download full-text PDF

Source
http://dx.doi.org/10.1002/micr.31160DOI Listing

Publication Analysis

Top Keywords

head neck
12
neck reconstruction
12
microvascular anastomosis
8
anastomosis head
8
coupling devices
8
ets anastomosis
8
anastomosis time
8
coupling device
8
anastomosis
6
comparison coupling
4