Article Abstract

Background: The first clinical trials to assess the feasibility of FLASH radiotherapy in humans have started (FAST-01, FAST-02) and more trials are foreseen. To increase comparability between trials it is important to assure treatment quality and therefore establish a standard for machine quality assurance (QA). Currently, the AAPM TG-224 report is considered as the standard on machine QA for proton therapy, however, it was not intended to be used for ultra-high dose rate (UHDR) proton beams, which have gained interest due to the observation of the FLASH effect.

Purpose: The aim of this study is to find consensus on practical guidelines on machine QA for UHDR proton beams in transmission mode in terms of which QA is required, how they should be done, which detectors are suitable for UHDR machine QA, and what tolerance limits should be applied.

Methods: A risk assessment to determine the gaps in the current standard for machine QA was performed by an international group of medical physicists. Based on that, practical guidelines on how to perform machine QA for UHDR proton beams were proposed.

Results: The risk assessment clearly identified the need for additional guidance on temporal dosimetry, addressing dose rate (constancy), dose spillage, and scanning speed. In addition, several minor changes from AAPM TG-224 were identified; define required dose rate levels, the use of clinically relevant dose levels, and the use of adapted beam settings to minimize activation of detector and phantom materials or to avoid saturation effects of specific detectors. The final report was created based on discussions and consensus.

Conclusions: Consensus was reached on what QA is required for UHDR scanning proton beams in transmission mode for isochronous cyclotron-based systems and how they should be performed. However, the group discussions also showed that there is a lack of high temporal resolution detectors and sufficient QA data to set appropriate limits for some of the proposed QA procedures.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.16854DOI Listing

Publication Analysis

Top Keywords

dose rate
16
proton beams
16
transmission mode
12
standard machine
12
uhdr proton
12
isochronous cyclotron-based
8
cyclotron-based systems
8
ultra-high dose
8
scanning proton
8
proton therapy
8

Similar Publications

The purpose of this study was to investigate the efficacy and safety of add-on metformin treatment in persons with active epilepsy (a-PWE). This is a single-centric, double-blind, placebo-controlled trial randomised a-PWE (1:1) to receive either metformin (extended-release 500 mg) or matching placebo for 6 months along with background antiseizure medications. Primary outcome was percentage change in seizure frequency/month, and secondary outcomes were 50% responder rate, serum mTOR expression, and serum total antioxidant capacity (TAC), body composition analysis, quality of life (QOL), and safety assessment.

View Article and Find Full Text PDF

A Monte Carlo Method for Estimating Secondary Photon Yields from Beta-emitting Radionuclides Concentrated in Environmental Soil.

Health Phys

September 2025

Nuclear and Radiological Engineering and Medical Physics Programs, George W. Woodruff School of Mechanical Engineering, Georgia Institute of Technology, Atlanta, GA.

External exposure due to secondary photons (predominantly bremsstrahlung) generated from electron source emissions in environmental soil are of concern due to their ability to deposit significant amounts of ionizing energy to organs and tissues within the body. The "condensed history method" employed in many modern Monte Carlo (MC) codes may be used to simulate secondary photon yields (given as photons per beta decay) arising from electron source emissions with relatively few assumptions regarding the secondary photon spatial, energy, and angular dependencies. These yields may in turn be used to derive protection quantities such as secondary photon effective dose rate (DR) and risk coefficients for a variety of idealized external exposure scenarios.

View Article and Find Full Text PDF

Background: While pegylated Escherichia coli asparaginase (PEG) is an integral component of leukemia and lymphoma treatment, hypersensitivity reactions (HSR) remain a common adverse event, often resulting in adjustments to the treatment regimen, increasing the burden on patients and families. HSR to asparaginase often indicates a transition to Erwinia asparaginase (ERW), which requires patients to return to the hospital 6 times for subcutaneous injections to replace one dose of IV PEG. Previous trials have demonstrated rates of HSR to pegylated E.

View Article and Find Full Text PDF

Purpose: To study the efficacy and safety of pro re nata regimen of brolucizumab, without loading dose, in treatment-naive patients with neovascular age-related macular degeneration (nAMD).

Case Series: Retrospective, observational study. We included all consecutive patients diagnosed with treatment- naïve nAMD undergoing Brolucizumab in Humanitas eye clinic, Turin, Italy between April 2022 and May 2023.

View Article and Find Full Text PDF

Background: Thrombotic thrombocytopenic purpura (TTP) is a life-threatening hematologic emergency caused by ADAMTS13 deficiency, leading to microvascular thrombosis, haemolytic anaemia, thrombocytopenia, and end-organ damage. Neurological symptoms occur in up to 90% of cases and are frequently misdiagnosed as stroke. Prompt recognition and treatment reduce the mortality rate from over 90% to 10-20%.

View Article and Find Full Text PDF