Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background: Our purpose was to evaluate a custom reverse total shoulder arthroplasty glenoid baseplate for severe glenoid deficiency, emphasizing the challenges with this approach, including short-term clinical and radiographic outcomes and complications.

Methods: This was a single-institution, retrospective series of 29 patients between January 2017 and December 2022 for whom a custom glenoid component was created for extensive glenoid bone loss. Patients were evaluated preoperatively and at intervals for up to 5 years. All received preoperative physical examinations, plain radiographs, and computed tomography (CT). Intra- and postoperative complications are reported.

Results: Of 29 patients, delays resulted in only undergoing surgery, and in three of those, the implant did not match the glenoid. For those three, the time from CT scan to implantation averaged 7.6 months (range, 6.1-10.7 months), compared with 5.5 months (range, 2-8.6 months) for those whose implants fit. In patients with at least 2-year follow-up (n=9), no failures occurred. Significant improvements were observed in all patient-reported outcome measures in those nine patients (American Shoulder and Elbow Score, P<0.01; Simple Shoulder Test, P=0.02; Single Assessment Numeric Evaluation, P<0.01; Western Ontario Osteoarthritis of the Shoulder Index, P<0.01). Range of motion improved for forward flexion and abduction (P=0.03 for both) and internal rotation up the back (P=0.02). Pain and satisfaction also improved (P<0.01 for both).

Conclusions: Prolonged time (>6 months) from CT scan to device implantation resulted in bone loss that rendered the implants unusable. Satisfactory short-term radiographic and clinical follow-up can be achieved with a well-fitting device. Level of evidence: III.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10698128PMC
http://dx.doi.org/10.5397/cise.2023.00563DOI Listing

Publication Analysis

Top Keywords

custom glenoid
8
reverse total
8
total shoulder
8
shoulder arthroplasty
8
months range
8
glenoid
5
patients
5
glenoid components
4
components reverse
4
arthroplasty background
4

Similar Publications

Is Acromial Morphology Different in Patients with Eccentric Glenohumeral Osteoarthritis?

J Shoulder Elbow Surg

August 2025

Department of Orthopaedic Surgery and Department of Physical Therapy, Washington University, St. Louis, MO, USA. Electronic address:

Background: Acromial morphology has been implicated as a potential contributor to eccentric glenohumeral osteoarthritis (GHOA), leading to the development of novel procedures including scapular spine corrective osteotomies. However, there remains a substantial gap in knowledge on the relationship between acromial morphology and eccentric GHOA. This study utilized a comprehensive three-dimensional (3D) semi-automated analysis of acromial morphology to assess its association with eccentric GHOA patterns.

View Article and Find Full Text PDF

» A thorough understanding of glenoid deformity is essential to anticipate intraoperative needs and manage bony deficiencies. Management strategies include eccentric reaming, alternate scapular line baseplate orientation, bone grafting, or metallic augmentation as either "off-the-shelf" or custom baseplates. Accurate implantation of the glenoid component in the presence of glenoid bone loss (GBL) may also benefit from shoulderspecific guides, navigation, or robotic assistance.

View Article and Find Full Text PDF

Introduction: Chronic protracted mandibular dislocation (CPMD) in elderly patients poses unique treatment challenges due to age-related anatomical changes. This report details the successful management of bilateral CPMD in a geriatric patient.

Case Report: A 69-year-old female presented with a 3-month history of inability to close her mouth following bilateral TMJ dislocation.

View Article and Find Full Text PDF

» Proximal humerus bone loss (PHBL) is a challenging issue in the setting of reverse total shoulder arthroplasty, particularly during revision surgeries. » PHBL can compromise the deltoid moment arm, soft tissue attachments, and stability of the humeral and glenoid components. » Causes of PHBL are multifactorial, and the degree of bone loss can vary widely among patients.

View Article and Find Full Text PDF

From resection to reconstruction: patient-specific temporomandibular joint replacement in patients with ameloblastomas - a report of three cases.

J Korean Assoc Oral Maxillofac Surg

June 2025

Department of Oral and Maxillofacial Surgery, JSS Dental College and Hospital, JSS Academy of Higher Education and Research, Mysuru, India.

Ameloblastomas involving the temporomandibular joint (TMJ) present unique reconstructive challenges. This case series introduces patient-specific TMJ prostheses as an innovative solution for complex mandibular defects, addressing critical limitations in traditional reconstruction methods by eliminating donor site morbidity and optimising functional outcomes. Three patients presented with follicular ameloblastomas in the posterior mandible, experiencing significant tumour involvement affecting mandibular function and facial aesthetics.

View Article and Find Full Text PDF