Quality assessment of anterior segment OCT images: Development and validation of quality criteria.

Photodiagnosis Photodyn Ther

Population, Policy and Practice Department, UCL Great Ormond Street Institute of Child Health, London, United Kingdom; Great Ormond Street Hospital, London, United Kingdom; NIHR Great Ormond Street Hospital Biomedical Research, London, United Kingdom; NIHR Moorfields Biomedical Research Centre, Lond

Published: February 2024


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background: The utility of medical imaging is dependant on image quality. We aimed to develop and validate quality criteria for ocular anterior segment optical coherence tomography (AS-OCT) images.

Methods: We undertook a cross-sectional study using AS-OCT images from patients aged 6-16. A novel three-level grading system (good, limited or poor) was developed based on the presence of image artefact (categorised as lid, eyelash, cropping, glare, or movement artefact). Three independent experts graded 2825 images, with agreement assessed using confusion matrices and intraclass correlation coefficients (ICC) for each parameter.

Results: There was very good inter-grader IQA agreement assessing image quality with ICC 0.85 (95 %CI: 0.84-0.87). The most commonly occurring artefact was eyelash artefact (1008/2825 images, 36 %). Graders labelled 621/2825 (22 %) images as good and 384 (14 %) as poor. There was complete agreement at either end of the confusion matrix with no 'good' images labelled as 'poor' by other graders, and vice versa. Similarly, there was very good agreement when assessing presence of lash (0.96,0.94-0.98), movement (0.97,0.96-0.99), glare (0.82,0.80-0.84) and cropping (0.90,0.88-0.92).

Conclusions: The novel image quality assessment criteria (IQAC) described here have good interobserver agreement overall, and excellent agreement on the differentiation between 'good' and 'poor' quality images. The large proportion of images graded as 'limited' suggests the need for refine this classification, using the specific IQAC features, for which we also report high interobserver agreement. These findings support the future potential for wider clinical and community care implementation of AS-OCT for the diagnosis and monitoring of ocular disease.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.pdpdt.2023.103886DOI Listing

Publication Analysis

Top Keywords

image quality
12
quality assessment
8
anterior segment
8
images
8
quality criteria
8
agreement assessing
8
interobserver agreement
8
quality
7
agreement
7
good
5

Similar Publications

Clinical Role of the Noninvasive Abdominal Fetal ECG in the Detection and Monitoring of Fetal Tachycardia.

Circ Arrhythm Electrophysiol

September 2025

Department of Congenital Heart Disease, Evelina London Children's Hospital, United Kingdom (S. Chivers, T.V., V.Z., S.M., G.M., W.R., E.R., D.F.A.L., T.G.D., O.I.M., G.K.S., J.M.S.).

Background: Fetal tachycardias can cause adverse fetal outcomes including ventricular dysfunction, hydrops, and fetal demise. Postnatally, ECG is the gold standard, but, in fetal practice, echocardiography is used most frequently to diagnose and monitor fetal arrhythmias. Noninvasive extraction of the fetal ECG (fECG) may provide additional information about the electrophysiological mechanism and monitoring of intermittent arrhythmias.

View Article and Find Full Text PDF

In recent AI-driven disease diagnosis, the success of models has depended mainly on extensive data sets and advanced algorithms. However, creating traditional data sets for rare or emerging diseases presents significant challenges. To address this issue, this study introduces a direct-self-attention Wasserstein generative adversarial network (DSAWGAN) designed to improve diagnostic capabilities in infectious diseases with limited data availability.

View Article and Find Full Text PDF

Few-shot learning for highly accelerated 3D time-of-flight MRA reconstruction.

Magn Reson Med

September 2025

Centre for Integrative Neuroimaging, FMRIB Division, Nuffield Department of Clinical Neurosciences, University of Oxford, Oxford, UK.

Purpose: To develop a deep learning-based reconstruction method for highly accelerated 3D time-of-flight MRA (TOF-MRA) that achieves high-quality reconstruction with robust generalization using extremely limited acquired raw data, addressing the challenge of time-consuming acquisition of high-resolution, whole-head angiograms.

Methods: A novel few-shot learning-based reconstruction framework is proposed, featuring a 3D variational network specifically designed for 3D TOF-MRA that is pre-trained on simulated complex-valued, multi-coil raw k-space datasets synthesized from diverse open-source magnitude images and fine-tuned using only two single-slab experimentally acquired datasets. The proposed approach was evaluated against existing methods on acquired retrospectively undersampled in vivo k-space data from five healthy volunteers and on prospectively undersampled data from two additional subjects.

View Article and Find Full Text PDF

Objective: To improve B field homogeneity in prostate MR imaging and spectroscopy using a custom-designed 16-channel external local shim coil array.

Methods: In vivo prostate imaging was performed in seven healthy volunteers (mean age: 40.7 years) without bowel preparation.

View Article and Find Full Text PDF

Purpose: Targeted therapy with lenvatinib is a preferred option for advanced hepatocellular carcinoma, however, predicting its efficacy remains challenging. This study aimed to build a nomogram integrating clinicoradiological indicators and radiomics features to predict the response to lenvatinib in patients with hepatocellular carcinoma.

Methods: This study included 211 patients with hepatocellular carcinoma from two centers, who were allocated into the training (107 patients), internal test (46 patients) and external test set(58 patients).

View Article and Find Full Text PDF