Article Abstract

Purpose: The purpose of this study was to compare the performance of Precise IQ Engine (PIQE) and Advanced intelligent Clear-IQ Engine (AiCE) algorithms on image-quality according to the dose level in a cardiac computed tomography (CT) protocol.

Materials And Methods: Acquisitions were performed using the CT ACR 464 phantom at three dose levels (volume CT dose indexes: 7.1/5.2/3.1 mGy) using a prospective cardiac CT protocol. Raw data were reconstructed using the three levels of AiCE and PIQE (Mild, Standard and Strong). The noise power spectrum (NPS) and task-based transfer function (TTF) for bone and acrylic inserts were computed. The detectability index (d') was computed to model the detectability of the coronary lumen (350 Hounsfield units and 4-mm diameter) and non-calcified plaque (40 Hounsfield units and 2-mm diameter).

Results: Noise magnitude values were lower with PIQE than with AiCE (-13.4 ± 6.0 [standard deviation (SD)] % for Mild, -20.4 ± 4.0 [SD] % for Standard and -32.6 ± 2.6 [SD] % for Strong levels). The average NPS spatial frequencies shifted towards higher frequencies with PIQE than with AiCE (21.9 ± 3.5 [SD] % for Mild, 20.1 ± 3.0 [SD] % for Standard and 12.5 ± 3.5 [SD] % for Strong levels). The TTF values at fifty percent (f) values shifted towards higher frequencies with PIQE than with AiCE for acrylic inserts but, for bone inserts, f values were found to be close. Whatever the dose and DLR level, d' values of both simulated cardiac lesions were higher with PIQE than with AiCE. For the simulated coronary lumen, d' values were better by 35.1 ± 9.3 (SD) % on average for all dose levels for Mild, 43.2 ± 5.0 (SD) % for Standard, and 62.6 ± 1.2 (SD) % for Strong levels.

Conclusion: Compared to AiCE, PIQE reduced noise, improved spatial resolution, noise texture and detectability of simulated cardiac lesions. PIQE seems to have a greater potential for dose reduction in cardiac CT acquisition.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.diii.2023.10.004DOI Listing

Publication Analysis

Top Keywords

piqe aice
16
piqe
8
dose levels
8
aice piqe
8
acrylic inserts
8
coronary lumen
8
hounsfield units
8
[sd] standard
8
[sd] strong
8
strong levels
8

Similar Publications

Purpose: The purpose of this study was to investigate whether the high-precision magnetic resonance (MR) sequence using modified Fast 3D mode wheel and Precise IQ Engine (PIQE), that was collected in a wheel shape with sequential data filling in the k-space in the phase encode-slice encode plane, is feasible for breath-hold (BH) three-dimensional (3D) T1-weighted imaging of the hepatobiliary phase (HBP) of gadoxetic acid-enhanced MRI in comparison to the compressed sensing (CS) sequence using Advanced Intelligent Clear-IQ Engine (AiCE).

Methods: This retrospective study included 54 patients with focal hepatic lesions who underwent dynamic contrast-enhanced MRI. Both standard HBP images using CS with AiCE and high-precision HBP images using modified Fast 3D mode wheel and PIQE were obtained.

View Article and Find Full Text PDF

Purpose: The purpose of this study was to compare the performance of Precise IQ Engine (PIQE) and Advanced intelligent Clear-IQ Engine (AiCE) algorithms on image-quality according to the dose level in a cardiac computed tomography (CT) protocol.

Materials And Methods: Acquisitions were performed using the CT ACR 464 phantom at three dose levels (volume CT dose indexes: 7.1/5.

View Article and Find Full Text PDF