Severity: Warning
Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 197
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3165
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 317
Function: require_once
98%
921
2 minutes
20
Rationale: Infrared matrix-assisted laser desorption electrospray ionization (IR-MALDESI) utilizes a 2970 nm mid-IR laser to desorb samples with depth resolutions (Z) on the order of micrometers. Conventionally, 5-20 μm thick tissue sections are used to characterize different applications of the IR-MALDESI source, but an optimal thickness has not been systematically investigated.
Methods: Mouse liver was sectioned to various thicknesses and analyzed using IR-MALDESI mass spectrometry imaging (MSI). Height profiles of tissue sections of various cryosectioned thicknesses were acquired to affirm tissue thickness. Tissue sections of each thickness were measured using a Keyence microscope. Paraffin wax was cryosectioned, mounted on microscope slides, and measured using a chromatic confocal sensor system to determine the cryostat sectioning accuracy.
Results: Analyzing sectioned tissues at higher thickness (>10 μm) leads to lower ion abundance, a decrease in signal over long analysis times, and more frequent instrument cleaning. Additionally, increasing tissue thickness above the optimum (7 μm) does not result in a significant increase in lipid annotations.
Conclusions: This work defines an optimal sample thickness for IR-MALDESI-MSI and demonstrates the utility of optimizing tissue thickness for MSI platforms of comparable Z resolution.