Severity: Warning
Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 197
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3165
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 317
Function: require_once
98%
921
2 minutes
20
Background: Real-world clinical data on the adverse events related to the use of cryoballoon catheter for pulmonary vein isolation remains limited.
Objective: To report and describe the adverse events related to the use of Artic Front cryoballoon catheters (Arctic Front, Arctic Front Advance, and Arctic Front Advance Pro) reported in the Food and Drug Administration's (FDA) Manufacturers and User Defined Experience (MAUDE) database.
Methods: We reviewed all the adverse events reported to the FDA MAUDE database over a 10.7-year study period from January 01, 2011 to September 31, 2021. All events were independently reviewed by two physicians.
Results: During the study period, a total of 320 procedural-related adverse events reported in the MAUDE database were identified. The most common adverse event was transient or persistent phrenic nerve palsy (PNP), accounting for 48% of all events. This was followed by cardiac perforation (15%), pulmonary vein stenosis (8%), transient ischemic attack or stroke (6%), vascular injury (4%), transient or persistent ST-elevation myocardial infarction (3%), hemoptysis (2%), pericarditis (2%), and esophageal ulcer or fistula (1%). There were six reported intra-procedural death events as a result of cardiac perforation.
Conclusion: The two most common procedural adverse events associated with cryoballoon ablation were PNP and cardiac perforation. All cases of procedural mortality were due to cardiac perforation.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10549805 | PMC |
http://dx.doi.org/10.1002/joa3.12898 | DOI Listing |