Article Abstract

Background: Balloon guide catheters (BGCs) have not been widely adopted, possibly due to the incompatibility of past-generation BGCs with large-bore intermediate catheters. The next-generation BGC is compatible with large-bore catheters. We compared outcomes of thrombectomy cases using BGCs versus conventional guide catheters.

Methods: We conducted a retrospective study of 110 thrombectomy cases using BGCs (n=55) and non-BGCs (n=55). Sixty consecutive thrombectomy cases in whom the BOBBY BGC was used at a single institution between February 2021 and March 2022 were identified. Of these, 55 BGC cases were 1:1 matched with non-BGC cases by proceduralists, age, gender, stent retriever + aspiration device versus aspiration-only, and site of occlusion. First-pass effect was defined as Thrombolysis In Cerebral Infarction 2b or higher with a single pass.

Results: The BGC and non-BGC cohorts had similar mean age (67.2 vs 68.9 years), gender distribution (43.6% vs 47.3% women), median initial National Institutes of Health Stroke Scale score (14 vs 15), and median pretreatment ischemic core volumes (12 mL vs 11.5 mL). BGC and non-BGC cases had similar rates of single pass (60.0% vs 54.6%), first-pass effect (58.2% vs 49.1%), and complications (1.8% vs 9.1%). In aspiration-only cases, the BGC cohort had a significantly higher rate of first-pass effect (100% vs 50.0%, p=0.01). BGC was associated with a higher likelihood of achieving a modified Rankin Scale score of 2 at discharge (OR 7.76, p=0.02). No additional procedural time was required for BGC cases (46.7 vs 48.2 min).

Conclusion: BGCs may be safely adopted with comparable procedural efficacy, benefits to aspiration-only techniques, and earlier functional improvement compared with conventional guide catheters.

Download full-text PDF

Source
http://dx.doi.org/10.1136/jnis-2023-020635DOI Listing

Publication Analysis

Top Keywords

guide catheters
12
thrombectomy cases
12
balloon guide
8
bgc
8
cases
8
cases bgcs
8
conventional guide
8
bgc cases
8
non-bgc cases
8
bgc non-bgc
8

Similar Publications

Guide Tip Damage Due to Rotablation.

JACC Case Rep

September 2025

Department of Cardiovascular Medicine, Okayama University Faculty of Medicine, Dentistry and Pharmaceutical Sciences, Okayama, Japan.

Background: The rotational atherectomy system can effectively debulk calcified coronary lesions. However, rare complications specific to that system have been reported.

Case Summary: A 77-year-old man with a heavily calcified lesion in the right coronary artery (RCA) ostium underwent percutaneous coronary intervention in an 8-F system.

View Article and Find Full Text PDF

Background: Diabetic foot ulcers (DFU) are a prevalent complication of diabetes, leading to significant morbidity, mortality, and amputation rates. Chronic non-healing DFU often result from peripheral neuropathy, microvascular issues, and infection, with poor blood and oxygen supply being critical factors in delayed healing. The development of new treatments to promote blood supply and accelerate ulcer healing is a significant area of research for DFU management.

View Article and Find Full Text PDF

Background: Optical coherence tomography (OCT) with artificial intelligence (AI) has been developed.

Aims: The study aimed to evaluate the differences between AI-quantified and visual assessments.

Methods: Patients scheduled for OCT-guided percutaneous coronary intervention between September 2021 and October 2022 were included.

View Article and Find Full Text PDF

Introduction: Wave speed (WS) mapping, enabled by omnipolar technology, allows for real-time visualization of local conduction velocity (CV). Its utility in ventricular tachycardia (VT) ablation has not been fully characterized.

Methods And Results: We describe a case series of patients undergoing VT ablation in which WS mapping was applied alongside established techniques such as peak frequency (PF) mapping and isochronal late activation mapping (ILAM).

View Article and Find Full Text PDF

Traditionally, clinical devices are designed, tested and improved through lengthy and expensive laboratory experiments and clinical trials [1]. More recently, computational methods have allowed for rapid testing, speeding up the design process and enabling far more complete searches of design space. While computational models cannot fully capture the complexities of biological systems, they provide valuable insights into crucial underlying mechanisms, such as the effects of fluid-structure interactions (FSIs).

View Article and Find Full Text PDF