Article Abstract

Background: This study aims to evaluate the trueness of the DI-60 Digital Cell Imaging Analyzer on Wright-stained samples with a focus on prevalence-dependent quality indicators for differential blood counts requested from non-hematology wards.

Methods: Two hundred and ninety-nine samples were included into this performance evaluation study at the Department of Laboratory Medicine, Medical University of Vienna. The following aspects were verified: (a) the reliability of automatedly pre-classified differential counts, (b) the concordance of DI-60 counts with manual-microscopic differential counts and (c) the agreement of DI-60 and manual-microscopic results regarding clinically relevant findings.

Results: 82.3% of all leukocytes were correctly pre-classified. Cell categories with a low prevalence (eosinophils, basophils, progenitors/precursors) in non-hematological patients presented with a low positive predictive value (PPV), indicating a high frequency of false positives. Comparisons between visually adjusted results of the DI-60 and manual-microscopic differential counts revealed a good concordance for neutrophil and lymphocyte counts. Besides the detection of precursors/progenitors and normoblasts, no relevant systemic errors were detected. However, due to their low prevalence and technical aspects, the detection of basophilia, monocytosis or the presence of precursors/progenitors showed comparably low accuracies (error rates of 7.4%-24.1%).

Conclusion: The DI-60 system works well for Wright-stained samples collected in the non-hematology ward. Due to the varying prevalence of cell categories found in peripheral blood, a low PPV can be expected with automatic assignment for those cells with low prevalence (e.g., basophils, eosinophils, precursor and progenitor cells, plasma cells). If the pre-test probability of these conditions is increased, manual microscopic processing may be recommended.

Download full-text PDF

Source
http://dx.doi.org/10.1111/ijlh.14179DOI Listing

Publication Analysis

Top Keywords

wright-stained samples
12
differential counts
12
low prevalence
12
di-60 digital
8
analyzer wright-stained
8
samples focus
8
focus prevalence-dependent
8
prevalence-dependent quality
8
quality indicators
8
manual-microscopic differential
8

Similar Publications

Background: This study aims to evaluate the trueness of the DI-60 Digital Cell Imaging Analyzer on Wright-stained samples with a focus on prevalence-dependent quality indicators for differential blood counts requested from non-hematology wards.

Methods: Two hundred and ninety-nine samples were included into this performance evaluation study at the Department of Laboratory Medicine, Medical University of Vienna. The following aspects were verified: (a) the reliability of automatedly pre-classified differential counts, (b) the concordance of DI-60 counts with manual-microscopic differential counts and (c) the agreement of DI-60 and manual-microscopic results regarding clinically relevant findings.

View Article and Find Full Text PDF

Background: The international consensus group suggested criteria for action following automated complete blood count and white blood cell differential analysis. These criteria were set based on data from laboratories of developed countries. It is highly important to validate the criteria in developing countries where infectious diseases are still rampant that can affect blood cell count and morphology.

View Article and Find Full Text PDF

Optimizing rat and human blood cells sampling for in silico morphometric analysis.

Acta Histochem

August 2022

Facultad de Química Farmacéutica Biológica, Universidad Veracruzana, Xalapa, Veracruz, Mexico. Electronic address:

Measurements of Morphometric Parameters of the Blood Cells (MPBC) are key for the diagnosis of both mental and metabolic diseases. Several manual approaches or computational methodologies are useful to provide reliable clinical diagnosis. The sample processing and data analysis is relevant, however the sample handling on the pre-analytical phase remains scarcely evaluated.

View Article and Find Full Text PDF

Southern giant pouched rats () are muroid rodents native to subSaharan Africa. They are increasingly used as service animals because of their keen sense of smell and are primarily known for clearing minefields in Africa. The objectives of this study were to determine hematologic and biochemical reference intervals from clinically healthy wild-caught captive adult rats, to describe the cytochemical staining reactions of peripheral blood leukocytes, and to document urinalysis findings.

View Article and Find Full Text PDF

Rationale: On Wright-stained sputum cytospins, eosinophil differential of ≥ 1.2% is considered abnormal, and ≥ 2.3% identifies an eosinophilic endotype.

View Article and Find Full Text PDF