Article Abstract

Background: Recent applications of high-resolution peripheral quantitative computed tomography (HR-pQCT) have demonstrated that changes in local bone remodelling can be quantified in vivo using longitudinal three-dimensional image registration. However, certain emerging applications, such as fracture healing and joint analysis, require larger multi-stack scan regions that can result in stack shift image artifacts. These artifacts can be detrimental to the accurate alignment of the bone structure across multiple timepoints. The purpose of this study was to establish a multi-stack registration protocol for evaluating longitudinal HR-pQCT images and to assess the accuracy and precision error in comparison with measures obtained using previously established three-dimensional longitudinal registration.

Methods: Three same day multi-stack HR-pQCT scans of the radius (2 stacks in length) and tibia (3 stacks in length) were obtained from 39 healthy individuals who participated in a previous reproducibility study. A fully automated multi-stack registration algorithm was developed to re-align stacks within a scan by leveraging slight offsets between longitudinal scans. Stack shift severity before and after registration was quantified using a newly proposed stack-shift severity score. The false discovery rate for bone remodelling events and precision error of bone morphology and micro-finite element analysis parameters were compared between longitudinally registered scans with and without the addition of multi-stack registration.

Results: Most scans (82 %) improved in stack alignment or maintained the lowest stack shift severity score when multi-stack registration was implemented. The false discovery rate of bone remodelling events significantly decreased after multi-stack registration, resulting in median false detection of bone formation and resorption fractions between 3.2 to 7.5 % at the radius and 3.4 to 5.3 % at the tibia. Further, precision error was significantly reduced or remained unchanged in all standard bone morphology and micro-finite element analysis parameters, except for total and trabecular cross-sectional areas.

Conclusion: Multi-stack registration is an effective strategy for accurately aligning multi-stack HR-pQCT scans without modification of the image acquisition protocol. The algorithm presented here is a viable approach for performing accurate morphological analysis on multi-stack HR-pQCT scans, particularly for advanced application investigating local bone remodelling in vivo.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.bone.2023.116893DOI Listing

Publication Analysis

Top Keywords

multi-stack registration
24
hr-pqct scans
16
bone remodelling
16
stack shift
12
precision error
12
multi-stack hr-pqct
12
multi-stack
11
accuracy precision
8
longitudinal hr-pqct
8
bone
8

Similar Publications

A 2D-registration algorithm for the correction of motion-induced misalignments of consecutive image stacks in multi-stack high-resolution peripheral quantitative CT scans.

Bone

August 2025

Department of Biomedical Engineering, Eindhoven University of Technology, Eindhoven, the Netherlands; Department of Orthopedic Surgery, Maastricht University Medical Center, Maastricht, the Netherlands. Electronic address:

Multi-stack imaging using high-resolution peripheral quantitative CT (HR-pQCT) can involve misalignments of consecutive image stacks ('stack shift') due to subject movement during scan acquisition. We developed a simple, 2D-registration algorithm for the correction of stack shifts in multi-stack HR-pQCT scans and investigated 1) the differences in standard HR-pQCT parameters and repeatability between before and after stack-shift correction; and 2) the correlation between the transformation needed for the stack-shift correction and corresponding difference in HR-pQCT parameters. The algorithm generates an artificial stack overlap of two slices, then rigidly registers the overlapping region (only in-plane translation allowed), and subsequently applies the resulting translation to the proximal stack.

View Article and Find Full Text PDF

Background: Recent applications of high-resolution peripheral quantitative computed tomography (HR-pQCT) have demonstrated that changes in local bone remodelling can be quantified in vivo using longitudinal three-dimensional image registration. However, certain emerging applications, such as fracture healing and joint analysis, require larger multi-stack scan regions that can result in stack shift image artifacts. These artifacts can be detrimental to the accurate alignment of the bone structure across multiple timepoints.

View Article and Find Full Text PDF

Background: Medical imaging plays an important role in determining the progression of joint damage in rheumatoid arthritis (RA). High resolution peripheral quantitative computed tomography (HR-pQCT) is a sensitive tool capable of evaluating bone microarchitecture and erosions, and 3D rigid image registration can be used to visualize and quantify bone remodeling over time. However, patient motion during image acquisition can cause a "stack shift" artifact resulting in loss of information and reducing the number of erosions that can be analyzed using HR-pQCT.

View Article and Find Full Text PDF

Accurate localization of anatomical landmarks is an important step in medical imaging, as it provides useful prior information for subsequent image analysis and acquisition methods. It is particularly useful for initialization of automatic image analysis tools (e.g.

View Article and Find Full Text PDF