98%
921
2 minutes
20
Background: Horizontal aorta (HA), defined by an aortic angulation (AA) ≥48°, is associated with worse outcomes particularly after self-expanding (SE) trans-catheter heart valve (THV) implantation. Although the SE ACURATE Neo THV demonstrated favorable procedural success rates in patients with HA, it remains associated with a non-negligible rate of moderate or greater paravalvular leak (PVL).
Objectives: Aim of the study was to assess the performance of ACURATE Neo2 in the setting of HA.
Methods: We performed a multicenter cohort analysis on patients with severe aortic valve stenosis and HA undergoing transcatheter aortic valve replacement (TAVR) with the Neo or Neo2 THV enrolled in the ITAL-neo registry. The primary endpoint was a composite of early safety and clinical efficacy at 30 days according to the Valve Academic Research Consortium-3 (VARC-3). Secondary endpoints included the occurrence of moderate or severe PVL and 90-day clinical outcomes.
Results: Among 900 patients included in the ITAL-neo registry, 407 exhibited HA; of these, 300 received a Neo THV and 107 a Neo2 THV. HA, irrespective of the THV implanted, emerged as an independent risk factor for developing ≥ moderate PVL. Technical and device success at 30-day follow-up was comparable between groups. However, Neo2 was associated with a significantly lower rate of ≥moderate PVL vs. Neo: (5% vs. 15%; p < 0.001), which was confirmed also at 90-day follow-up. Additionally, no correlation was found between ≥moderate PVL and AA in the Neo2 group, while PVL degree increased proportionally to the AA in the Neo cohort.
Conclusion: In patients with HA, the new generation Acurate Neo2 THV was associated with a comparable device success rate and a significantly lower rate of ≥moderate PVL, when compared with its predecessor.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.ijcard.2023.131236 | DOI Listing |
Multimed Man Cardiothorac Surg
September 2025
Department of Cardiothoracic Surgery, Royal Children’s Hospital, Melbourne, Australia
The patient had rheumatic heart disease, which resulted in severe aortic and mitral valve regurgitation. Repair of both valves was performed at 9 years of age. During surgery, the retracted aortic valve cusps required extension with bovine pericardial patches and suture reduction annuloplasty, and the mitral valve was repaired using a Cosgrove-Edwards (Edwards Lifesciences LLC, Irvine, CA) annuloplasty band.
View Article and Find Full Text PDFJ Thorac Cardiovasc Surg
September 2025
Division of Cardiac Surgery, Department of Surgery, Western University, London Health Science Centre, London, Ontario, Canada.
J Neuroendovasc Ther
August 2025
Department of Neurosurgery, Graduate School of Biomedical and Health Sciences, Hiroshima University, Hiroshima, Hiroshima, Japan.
Objective: Endovascular treatment of cerebrovascular diseases can be challenging in patients with difficult access routes. We describe a turn-over technique using a balloon guiding catheter (BGC) via the transfemoral approach to perform neuroendovascular treatment.
Case Presentation: An 80-year-old female with a recurrent right middle cerebral artery (MCA) aneurysm after coil embolization underwent successful stent-assisted coiling via the transfemoral approach.
J Geriatr Cardiol
August 2025
Department of Bachelor of Medicine, Bachelor of Surgery, Sindh Medical College, Jinnah Sindh Medical University, Karachi, Pakistan.
Cureus
August 2025
Acute Internal Medicine, University Hospitals of North Midland, Royal Stoke University Hospital, Stoke-on-Trent, GBR.
Infective endocarditis is a potentially fatal condition that can present with non-specific symptoms and rare hematologic manifestations, posing significant diagnostic challenges. We report a compelling case of a 67-year-old male with a history of type 2 diabetes, hypertension, and hyperlipidemia who sought medical attention for a five-month history of progressive iron deficiency anemia, accompanied by weight loss, fatigue, and vague constitutional symptoms. Initial extensive workup, including computed tomography of the thorax, abdomen, and pelvis, gastroscopy, colonoscopy, and transthoracic echocardiography, failed to identify an underlying cause.
View Article and Find Full Text PDF